(C-4) Logging Cycle
1) When [LOG TIME] is displayed, press both
[ AIR ADJ.] and [ POWER] buttons at the
same time.
--> The current logging cycle “_ _ SEC (seconds)”
will be indicated.
Press the [ AIR ADJ.] button to select the cycle.
--> The value (second) will change by pressing the
button.
Configurable Logging Cycle (seconds)
0.5
1, 2, 3, 4, 5, 6, 7, 8, 9, 10
20, 30, 40, 50, 60(1min)
120, 180, 240, 300, 360, 420, 480, 540, 600(10min.)
1200, 1800, 2400, 3000, 3600(60min.)
2) To save the setting, press both [ AIR ADJ.] and
[ POWER] buttons at the same time.
--> “SAVE OK?” will be displayed.
3) Press both [ AIR ADJ.] and [ POWER] buttons at
the same time for OK.
--> Returns to [LOG TIME] screen after two beep sounds
4) To continue each setting, press the [ AIR ADJ.] button
and select next setting.
5) To quit the user mode, keep pressing the [ POWER]
button until it returns to the <Gas Concentration Display>.
MENU
(Logging cycle is
set to 60 seconds)
Blinking
•
The factory default of logging cycle is 10 sec.
•
The logging period is about 40 hours when the logging cycle is 10 sec.
NOTE
MENU
MENU
MENU
▲
How to reverse the display order of selection list
Press and hold both [ AIR ADJ.] and [ POWER]
buttons at the same time to reverse the display order of
selection list. After a beep sound, an arrow sign in the bottom
right corner of the LCD will flip vertical.
The same operation can reverse the display order again.
The display order will return to the default ( up arrow) when
the power is turned OFF.
NOTE
MENU
(Flipped down
arrow )
▲
▲
▲
▲
▲
(Current logging
cycle is 10)
Blinking
<Gas Concentration Display>
- 25 -
(C-2) Stop Logging
1) When [LOG ON] is displayed, press both [ AIR ADJ.]
and [ POWER] buttons at the same time.
--> “LOG OFF” will be indicated.
2) To set OFF (stop), press both [ AIR ADJ.] and
[ POWER] buttons at the same time.
--> “LOG OFF” will be indicated.
3) To continue the logging mode, press the [ AIR ADJ.]
button and select next setting.
4) To quit user mode, keep pressing the [ POWER]
button until it returns to the <Gas Concentration
Display>.
(C-3) Delete Logging Data
1) When [LOG DEL] is displayed, press both [ AIR ADJ.]
and [ POWER] buttons at the same time.
--> “DEL OK?” will be displayed.
2) Press both [ AIR ADJ.] and [ POWER] buttons at
the same time for OK.
--> Delete is executed with two beep sounds, and returns
to [LOG DEL] screen.
3) To continue logging mode, press the [ AIR ADJ.]
button and select next setting.
4) To quit user mode, keep pressing the [ POWER]
button until it returns to the <Gas Concentration
Display>.
MENU
MENU
(Logging is ON)
Blinking
Blinking
Blinking
MENU
(Data logging is OFF)
<Gas Concentration Display>
All the data will be erased after the logging data is deleted.
NOTE
MENU
MENU
MENU
▲
▲
▲
▲
▲
▲
<Gas Concentration Display>