- 48 -
11. Specifications
Detection principle
Electrochemical with/without pyrolyzer, Hot-wire semiconductor,
Catalytic, or Galvanic cell
Sampling method
Extractive pump (500mL/min. auto-controlled)
Sampling tube
PTFE, Outside and inside dia.: 6mm and 4mm (or 1/4 in. and 11/64 in.)
*1
Piping length: < 20m
*2
Target gas
As per delivery specifications
Detection range
As per delivery specifications
Gas concentration
display
Four-digit LEDs on LCD display, with unit of measurement
20 segment bar graph
Alarm set value
As per delivery specifications
Alarm accuracy
Combustible gas:
25% of alarm set value under identical conditions.
Toxic gas:
30% of alarm set value under identical conditions.
Low oxygen:
1vol% of alarm set value under identical conditions.
Alarm delay
Combustible gas:
≤ 30 seconds with a gas concentration that is 1.6 times
higher than the alarm set concentration.
Toxic gas:
≤ 60 seconds with a gas concentration that is 1.6 times higher
than the alarm set concentration.
Low oxygen:
≤ 5 seconds until the reading reaches 18vol% with a 10vol%
concentration at 20
C
2
C.
*Excludes delay time caused by piping length and communication time.
Alarm indication
Gas alarm (2 stages)
Red LED flashes.
ALARM1and/or ALARM2 shown on display.
Filter element replacement notification
Flow rate icon slowly spins on display.
Low flow rate alarm (with 10 second delay time)
Amber LED flashes.
FLOW shown on display. Flow rate icon stops.
Sensor fault/incorrect sensor installed
Amber LED flashes.
SENS. shown on display.
Broken pyrolyzer wire
Amber LED flashes.
CONV. shown on display.
External outputs
Digital signal: Ethernet100base-Tx(Modbus/TCP)
*3
Modbus/TCP (Max. number of connected units depends on the system
configuration.)
Max. transmission distance: 100m up to HUB
Baud rate: 100Mbps
Digital signal:
RS-485 (Modbus/RTU)
*3
Modbus-RTU (Max. number of connected units:32 units including master)
Method: 2-wire half duplex
Max. transmission distance: 1.2km up to the host device
Baud rate: 9600bps
* Selectable to Ethernet or RS-485 using DIP switches.
Gas concentration analog output signal
4-20mADC (common negative with power supply)
Output accuracy: ±0.5% of full scale
* 0.6mA or less in the event of a trouble alarm
*
300 Ω or less including wiring resistance