Filling/refilling the liquid
Use the syringe provided as a standard accessory with the equipment. Insert the syringe with double distilled water into one of pipes (any pipe)
and fill up to the level indicated on the panel. The equipment holds roughly 800ml of bi-distilled water. If overfilled the water will drain out from
the other pipe
Before closing the rear panel, verify the presence of the stoppers on the 2 small tubes.
Cleaning the filters
Failure to maintain these filters can impair the device's performance.
Do not cover the vents on the device with cloth or other materials. Leave at least 10 centimeters of space between the vents and
walls or obstacles. Failure to do so may compromise the efficiency of the device.
Do not cover any of the exhausting grids with clothes or other things and leave enough space (at least 10cm) from the rear grid and
the wall..
Cleaning of air filter
Unscrew the 4 screws shown in the photo. Clean up the filter with compressed air.
Replace the filter when it is very dirty; replace with a filter of the same type (Code RIC00041).
Replace regularly this filter because when full of dust and dirt may compromise the proper
functioning of the equipment.
Replacing the protective glass slide
Verify at least once in a month the state of the glass placed inside of the room (see photo).
To remove the door glass, unscrew the 2 screws indicated in the figure. In case the glass is
saturated of metal sprays or damaged, replace it with one new (Code ATE00018). 1 - Check
also the black paper (Code D2139193) cone inside the plastic light diffuser, to get the best
performance of the illumination this paper must be in good condition, replace it when
necessary.
If the glass is very dirt or damaged, the power of the equipment can go down until to 80% regarding the normal performances.
Checking alignment
•
Set the following parameters 1.0KW, 1.0mS 0Hz and minimum SPOT size.
•
Insert a piece of flat metal and fire a shot.
•
Check the location of the spot produced against the crosshair; if the spot intersects the crosshair in the center is not necessary to correct the
alignment.