the unit enters the relaxation cycle. When Trigger mode is disabled (by pressing
the intensity button) the stimulation builds over a 2-second period to the previously set
amplitude level.
Caution:
In trigger mode, where stimulation is held constant for several continuous
seconds, muscle fatigue could occur. This mode of operation may provide relief
only from muscle spasm.
16. AvivaTens XP has a load sense function that monitors the connection between the
lead and its electrode and between the electrode and the skin. When poor skin
contact is detected (Display 9):
• The amplitude bar of the channel being used will flash.
• The warning icon ( ) flashes on the display.
• An audible beep will emit from the unit.
• The treatment session timer pauses.
• The intensity value falls to zero and the up-arrow intensity button is deactivated.
When proper contact is restored, stimulation builds over a 2-second period to
the previously set intensity level.