NeuLog NUL-223 Скачать руководство пользователя страница 2

NEULOG DROP COUNTER LOGGER SENSOR GUIDE

 

 

 

Tablet, smart phone device 

 

Materials needed: 
 

 

NUL-223 Drop Counter Sensor 

 

WIFI-201 WIFI Module 

 

BAT-200 Battery Module 

 
Your drop counter sensor needs to be connected to a WIFI-201 module. The WIFI-201 
module will create a closed NeuLog WiFi network which will stream the NeuLog data 
to a device of your choosing. Once your device is wirelessly connected to the NeuLog 
network,  you  can  run  experiments  and  collect  data  through  a  browser  of  your 
choosing. 
 

Procedure: 

 

1.

 

Connect the drop counter sensor directly to the left side of a WIFI-201 (no wires 
required). 

2.

 

Connect a BAT-200 module to the right side of the WIFI-201 module. 

3.

 

Although not required, we recommend plugging the BAT-200 to an outlet using a 
USB to mini USB charger (such as a typical cell phone charger). The WIFI-201 
module  will  run  for  60-90  minutes  (depending  on  the  sensor)  without  being 
plugged in. 

4.

 

The WIFI-201 can be powered directly using a mini to USB cord and plugging it 
into your computer or a wall charger. Please note this function is only available 
on WIFI-201 modules with the USB icon on the cover. 

5.

 

For further WIFI-201 instructions or the WIFI-201 quick start guide please visit: 

www.NeuLog.com

Choose "Products" on the main menu, 

then choose "WiFi Communication module". 

6.

 

The WIFI-201 indicator lights will flash; take no action until the LED to the far left 
turns blue; this can take up to one minute. 

7.

 

Take your tablet or smart phone, go to the WiFi settings and select the NeuLog 
network (NeuLog0184 for example) which matches the ID found on the back of 
the WIFI-201 module (0184 for example). 

8.

 

Give your device about 20 seconds to connect to the WIFI-201. 

9.

 

Once  the  device  is  connected,  go  to  your  browser  and  type  the  website 

wifi201.com

 into the URL bar, then wait for 30-60 seconds. 

10.

 

You will see a “Control mode” icon in the application, click on it. 

11.

 

The  application  will  then  load  a  new  screen  and  begin  to  auto  detect  the 
sensor(s); this can take a minute. (You can stop the search when the sensor is 
found.) 

12.

 

If the application does not auto detect the sensor(s), select “Search for sensors”. 

13.

 

Once the sensor is found you will see a module box on the left side of the screen 
for the drop counter sensor. In this box, real time data will be displayed. 

14.

 

Click  on  the  “Module  setup”  button  located  on  the  drop counter  sensor  icon  in 
the module window to change the drop counter probe settings if need be. 

15.

 

Click on the “Experiment setup” icon to change the experiment settings if need 
be (Experiment duration for example). 

16.

 

To run an experiment and collect data, click the “On-Line experiment” icon, then 
click the "Run experiment" icon. 

17.

 

If you have a newer generation  WIFI-201 (with the USB icon on the cover) this 
can  be  used  as  a  USB-200,  meaning  you  can  tether  the  probes  directly  to  the 
computer using the WIFI-201. To engage operation in this fashion, press 3 times 
on  the  panel  button.  Repeating  this  operation  will  return  the  module  to  WiFi 
mode. 

 
 

Operation with NeuLog Graphic Display Module (viewer) 

 

Materials needed: 

 

NUL-223 Drop Counter Sensor 

 

VIEW-101 Graphic Display Module 

 

BAT-200 Battery Module 

 

Procedure: 

 
1. 

Connect the drop counter sensor to the left side of the viewer. 

2. 

Connect the BAT-200 module to the right side of the viewer. 

3. 

The VIEW-101 will auto detect the sensor and you will see it appear on the left 
side of the screen. 

4. 

Once the sensor appears it will be monitoring data in real time. 

5. 

To run an experiment and collect data press the run icon (little green person). 

 
 

Off-line experiments 

 
(Off-line  experiments  are  when  you  do  not  have  a  sensor  connected  directly  to  a 
computer, tablet, smartphone, or NeuLog viewer). 
 

Materials needed: 

 

NUL-223 Drop Counter Sensor 

 

BAT-200 Battery Module 

 

Содержание NUL-223

Страница 1: ... Quick start procedure PC or Mac Computer Materials needed NUL 223 Drop Counter Sensor USB 200 USB Module USB to mini USB cable included with the USB 200 module Your drop counter sensor needs to be connected to a USB 200 module The USB 200 module then connects to a computer via a USB to mini USB cable Please note that you cannot plug the drop counter sensor directly into the computer The browser b...

Страница 2: ...nto the URL bar then wait for 30 60 seconds 10 You will see a Control mode icon in the application click on it 11 The application will then load a new screen and begin to auto detect the sensor s this can take a minute You can stop the search when the sensor is found 12 If the application does not auto detect the sensor s select Search for sensors 13 Once the sensor is found you will see a module ...

Страница 3: ... Procedure 1 Connect the USB 200 to a PC Mac tablet or smart device 2 Connect the drop counter sensor to a USB 200 module they directly plug together 3 Open the NeuLog application 4 Click the Tools icon 5 Click Restore sensor s factory defaults on the menu Included with the sensor NeuLog NUL 223 Drop Counter Sensor Guide this document Durable plastic drop counter probe attached directly to the sen...

Страница 4: ...quid s opacity or color the infrared light bends away from the detector Each time there is a break in the infrared beam a drop is counted Maintenance and storage Never submerge the NeuLog plastic body in any liquid Do not allow liquid into the drop counter sensor s body and probe After use gently wipe away any foreign material from the drop counter sensor Store in a box at room temperature out of ...

Отзывы: