1.  INTRODUCTION 

This Micro Centrifuge offers versatility, simplicity, safety, and ease of usage in a 

single package. This product has been designed after studying centrifuges and 

its usage in numerous scenarios of the lab environment. 
This Micro centrifuge is specially suitable for microfiltration, cell separations, 

centrifugal tube sedimentation, and spin down applications. Please read this 

manual thoroughly before operating the centrifuge. 

2.  INTENDED USE 

This is used in various laboratories to separate particles from suspension or even 

macromolecules  from  solutions  according  to  their  density.  The  different 

biological substances that are usually separated by centrifugation are microbial 

cells, organelles, proteins, DNA, and RNA. 

NOTE: This user guide is intended to assist with the operation and care of the unit 

only and not for its repair. The customer should not attempt to service or repair the 

unit. ff repairs are required, please contact the supplier. 

3.  FEATURES 

• 8 position micro centrifuge tube rotor

• Digital display with set variable speed & timer

• Electronic safety brake on lid opening for user safety

• Magnetic rotorfor easytool free changing of rotor 

• Startsand stops in seconds

• Lid safety i nterlock 

• Smooth and silent, discreet running 

• Small footprint can be fit in hoods & small work places

• Interchangeable magnetic  rotors (MCT rotor & PCR rotor) 

• Low voltage power consuming device

4.  STANDARD ACCESSORIES 

8x 1.5/2ml rotor, PCR strip rotor

Individual adaptor for 0.5 ml tu bes, pack of 8

Individual adaptor for 0.1/ 0.2 ml tu bes, pack of 8

Power Adaptor: Input voltage: 100 - 240VAC, 50/60Hz

Output voltage: 24V :-:-: 1.5A 

Product User Manual, Warranty Card and Declaration Conformity

5.  TECHNICAL SPECIFICATIONS 

Motor Type 

DC Motor 

Rotor Capacity 

8 x 1.5 /2.0 ml Tube Rotor & 2 Place PCR 

Strip Rotor 

1000 - 6000 RPM 

Speed 

2000 x g (MAX RCF) 

Speed Accuracy 

± 100 RPM 

Time 

1 - 25 & Infinate min. 

Size (W X D X H) 

162 x 157 x 116 mm 

Weight 

1.1 kg 

Input Voltage 

24V ;;-; 0.62 A 

Power Consumption 

15W 

Содержание iFUGE M08VT

Страница 1: ...NEUATION TECHNOLOGIES j iFUGE MOSVT PRODUCT USER MANUAL ...

Страница 2: ...hould not attempt to service or repair the unit ffrepairsarerequired pleasecontactthesupplier 3 FEATURES 8positionmicrocentrifugetuberotor Digitaldisplaywithset variablespeed timer Electronicsafetybrakeonlid openingforusersafety Magneticrotorfor easytool freechangingofrotor Startsand stopsinseconds Lidsafetyinterlock Smoothandsilent discreetrunning Smallfootprintcanbe fitinhoods smallworkplaces In...

Страница 3: ...ge whileit sunderoperation Do not fill the tubes while they are in the rotor Liquid spillage may harm the unit Do not put hands in the rotor chamber unless the rotor has completely stopped Donotcentrifuge flammable explosiveor corrosive materials Do not use solvents or flammable media near this or other electrical equipment In case the unit is dropped from the table please check the casing for any...

Страница 4: ...trifuge 4 The following are the incorrect methodsof loading tubes in the centrifuge the rotor 5 The following are the incorrect methodsof loading tubes in the centrifuge the rotor 9 INTERCHANGEABLE ROTOR 8 Slot Magnetic Rotor 8 X0 2 0 4 0 5 1 5 2ml 10 CHANGINGTHE ROTOR This centrifuge feature magnetic rotors Thus changing the rotors can be done without any tool The rotors fit an the central shaft ...

Страница 5: ... defects in original material or workmanship This warranty does not cover any incidental or consequential damages commerciallossor any other damages fromthe useofthisproduct The warranty is invalidated by any non factory modification which will immediatelyterminate all liabilities on us for the productsor damages caused by its use The customer shall be responsible for the product or use of product...

Страница 6: ...ations based on EU Directive 2012 19 EU on waste electrical and electronic equipment WEEE According to these regulations any devices supplied after 13 06 05 in the business to business sphere to which this product is assigned may no longer be disposed off in municipal or domestic waste They are marked withthe followingsymboltoindicatethis As disposal regulations within the EU may vary from country...

Отзывы: