Data Report 

 

After power on, the device will immediately send a version packet report, and then send another   
report with the data of ORP value 

after 20 seconds

The device sends data according to the default configuration before any other configuring. 
 

MaxTime: Max Interval=3600s (60min) 
MinTime: The MinTime configuration is not available. But the software has restriction, MinTime   
must be configured a number greater than 0. 

 
Note: 

1.Report interval is based on the factory default. 
2.R718PA9 reports the ORP value. 

 

The device reported data parsing please refer to                                                                                                                                     
Netvox LoraWAN Application Command document and Netvox Lora Command Resolver   
http://www.netvox.com.cn:8888/page/index 

 

 

Report Con figuration 

 

(1)

 

Configure R718PA9 device parameter MaxTime = 1min 

          Downlink: 0157000A003C0000000000 
          Device Return: 
                        8157000000000000000000 (configuration success) 
                        8157010000000000000000 (configuration failure) 
 
(2)

 

Read R718PA9 device parameter 

          Downlink: 0257000000000000000000 
          Device Return: 
                        8257000A003C0000000000 (device current parameter) 

Description 

Device 

Cmd 

ID 

Device

Type 

NetvoxPayLoadData 

ConfigReport

Req 

R718PA9 

0x01 

0x57 

 

MinTime 

(2bytes Unit: s) 

MaxTime 

(2bytes Unit: s) 

Reserved 

(5Bytes,Fixed 0x00) 

ConfigReport

Rsp 

0x81 

Status 

(0x00_success) 

Reserved 

(8Bytes, Fixed 0x00) 

ReadConfig 

ReportReq 

0x02 

Reserved 

(9Bytes, Fixed 0x00) 

ReadConfig 

ReportRsp 

0x82 

MinTime 

(2bytes Unit: s) 

MaxTime 

(2bytes Unit: s) 

Reserved 

(5Bytes,Fixed 0x00) 

Содержание R718PA9

Страница 1: ...td This document contains prORPietary technical information which is the property of NETVOX Technology It shall be maintained in strict confidence and shall not be disclosed to other parties in whole...

Страница 2: ...1 Table of Content 1 Introduction 2 2 Appearance 2 3 Main Feature 3 4 Set Up Instruction 4 5 Data Report 5 6 Information about Battery Passivation 6 7 Important Maintenance Instruction 7...

Страница 3: ...mpared with other communication methods LoRa spread spectrum modulation method greatly increases to expand the communication distance Widely used in long distance low data wireless communications For...

Страница 4: ...d spectrum technology Configuring parameters and reading data via the third party software platforms and set alarms via SMS text and email optional Applicable to the third party platforms Actility Thi...

Страница 5: ...k Turn on the device to search the network The green indicator keeps on for 5 seconds success The green indicator remains off fail Had Joined the Network Not in the original setting Turn on the device...

Страница 6: ...ument and Netvox Lora Command Resolver http www netvox com cn 8888 page index Report Con figuration 1 Configure R718PA9 device parameter MaxTime 1min Downlink 0157000A003C0000000000 Device Return 8157...

Страница 7: ...ery passivation may also lead to voltage delay when the batteries are put into operation and our devices may not work correctly in this situation As a result please make sure to source batteries from...

Страница 8: ...e parts and electronic components Do not store in excessive heat place High temperatures can shorten the life of electronic devices destroy batteries and deform or melt some plastic parts Do not store...

Отзывы: