Install the sensor at the bottom of the container / tank:
1. Power on first to see if the sensor is working normally. The red light flashing on the sensor indicates that the sensor can work normally, as shown in
the figure.
2. Place the container or tank in a horizontal position, preferably the angle between the container or tank and the horizontal plane is less than 2 degrees.
3. Make sure the container / tank is full or at least 2/3 of the liquid.
4. Find the best detection point for installing the sensor (the bottom of the container should be kept clean).
A. Apply some coupling glue on the sensor surface.
b. Place the sensor on the bottom of the container / tank; draw a circle around the test point and gently move to find the correct test point.
How to find the correct detection point: Check the reported installation information. 0 means that it can be installed at this detection point; 1
means it cannot be installed at this detection point.
Note:
When selecting the correct detection point, pay attention to the following points:
a) Ensure that the working surface of the sensor is in a horizontal position, and the working surface is parallel to the liquid level in the tank; the
radiation axis of the sensor is perpendicular to the horizontal direction. As the picture shows in below:
b) The 1, 2, and 4 points in the above figure are not suitable for mounting sensors because they are too close to the tank wall or have obstacles. Point 5
is not suitable because it is not horizontal. Position 3 is suitable for mounting the sensor.
5. Clean the checkpoint and the sensor surface, and prepare to fix the sensor to the container / tank.
6. Apply AB glue to the surface of the sensor, stick the sensor to the inspection point of the container / tank, and keep it for a while until it is fixed.
7. If the installation is correct, the device reports the installation status in REPORT as 00, otherwise, it’s 01.