11
KEYPAD FUNCTIONS
[
]
Power ON/OFF button. At default settings („Auto Off“ is displayed) the unit will
power off automatically after 5 min non-use to save battery power. Pressing the
button again, this function will be turned off (“Auto Off” is no longer displayed).
[dB/dBm]
Pressing this key toggles between measuring method absolute level (dBm)
and attenuation (dB). Holding the key until „HELD“ is displayed (appx 7 sec)
switches to
µ
W mode.
[Zero]
Pressing the key will set the OPM to zero position. (see note below)
[Ref]
This key recalls the last reference value, which is stored in memory. To define
a new reference value press the key 2 times. When OPM is set to dB mode,
the difference in dB between reference value and current input signal will be
displayed.
[
λ
]
Pressing this key selects the required wavelength according to the optical fiber
and the light source connected to OPM.
[ Ext. Power Indicator]
The external power indicator LED is lit when an external AC
power adapter is connected.
NOTE
How to set OPM to zero position:
• Ensure that the dust cap fits tight at the optical fiber input in order to prevent it from
extraneous light.
• After unit is powered up, keep the „Zero“ button pressed until „SUCC“ is displayed.
This confirms the correct zero setting.
• If „ERR“ is displayed, the auto zeroing procedure was unsuccessful. Check again,
whether the dust cap fits tight to connector and perform zero setting again.
• If still unsuccessful, please contact customer service of NetPeppers GmbH for
assistance.