background image

User Manual                                    DigiMano 1000 

 

 

10 

 

Operating Instructions: 

1.  Turn the DigiMano 1000 ON and allow two minutes for the transducer to warm up 

and stabilize. 
 

2.  Select the desired engineering unit to be displayed. 

 

3.  With no pressure applied the display should read zero (0). If not, adjust the 

manual zero adjustment potentiometer. 
 

4.  Connect the DigiMano 1000 to the device to be tested or to a pressure vacuum 

line. 
 

5.  Read the pressure or vacuum in the measurement units that have been selected 

by the Unit Selector Switch. The negative sign (-) will be displayed when reading 
vacuum measurements. 
 

Precautions 

 

1.  Do not exceed the rated maximum pressure of the instrument, although a 

transient pressure of two times the rating will generally be tolerated. 

 

2.  Do not connect the instrument to corrosive gases or highly ionic fluids. 

 

3. 

Do not continue to use if the instrument is wet.

 

Trouble Shooting 

 

When inaccurate pressure vacuum readings are suspected, check the following items: 
 

1.  Install a fresh 9 Volt battery. 

 

2.  Turn the DigiMano 1000 on and verify that the display reads zero with no 

pressure applied to the instrument, 00+/- 1 count. 
 

3.  Adjust the zero adjustment potentiometer as necessary. 

 

4.  Connect the instrument to a pressure source, apply a known pressure, and verify 

the reading within the specifications of the DigiMano 1000 model, 0.25%+/- 1 
count. 

 
 
 
 
 
 
 

Содержание DigiMano 1000

Страница 1: ...DIGIMANO 1000 Netech Corporation 110 Toledo St Farmingdale NY 11735 http www NetechCorp us http www PressureMeter com DigiMano 1000 Pressure Vacuum Meter INSTRUCTION MANUAL...

Страница 2: ...s 4 Returns Credits 5 Repair and Recalibration 6 Models and Part Numbers 7 Optional Accessories 7 General Overview Introduction 8 Specifications 8 General Operation 9 Operating Instruction 10 Precauti...

Страница 3: ...without notice and without incurring any obligation Netech guarantees availability of service parts for 5 years after the manufacture of the unit is discontinued The warranty is void if you elect to h...

Страница 4: ...etech is ISO 9001 2015 Certified This instrument was thoroughly tested and inspected according to Netech s ISO 9001 2015 quality standards and test procedures and found to meet those specifications wh...

Страница 5: ...of purchase or products which are not in as new and resalable condition are not eligible for credit return and will be returned to the customer The Return Procedure see below must be followed to assur...

Страница 6: ...be clearly marked on the package along with a statement indicating the reason for return Repair and Calibration Products returned for repair or recalibration must obtain a RMA Return Material Authoriz...

Страница 7: ...200 2000CMH2O DigiMano 1000 2000 CmH2O Pressure Vacuum Meter Range of 19 99 PSI and 1999 CmH2O Accuracy 0 25 FSO 200 BP CAL KIT Kit includes DigiMano 1000 P N 200 2000PS 1999mmHg 19 99PSI Carrying cas...

Страница 8: ...It is powered by one 9 Volt alkaline battery Optionally an AC adapter may be used The DigiMano 1000 is CE marked and is shipped with a Certificate of Calibration traceable to the NIST The DigiMano 10...

Страница 9: ...3_ Digit LCD indicates the result of the measurement being made 2 On Off Switch Turns instrument on and off 3 Unit Selector Switch Selects desired units of measurement 4 Zero Adjust Manual zero adjust...

Страница 10: ...measurements Precautions 1 Do not exceed the rated maximum pressure of the instrument although a transient pressure of two times the rating will generally be tolerated 2 Do not connect the instrument...

Страница 11: ...t and fast turnaround Estimates for repair and recalibration are available upon request DigiMano 1000 contains NO USER SERVICEABLE PARTS and calibration service should be performed only by Netech Atte...

Страница 12: ...User Manual DigiMano 1000 12 Appendix This section is intentionally left blank for future use...

Отзывы: