Installation Guide
Part #:
70-0285
(See DSU-H Instructions for Use)
Questions or issues?
Let us help!
Call:
(201) 343-5202, Ext. 1
DSU-H In-Line (continued)
AVOID contact between the filter and cleaning or sanitizing agents.
DO NOT remove blister pack from filter until indicated to protect from
contamination.
While wearing gloves, HOLD the filter by the
transparent portion of the blister pack and tear
away the white backing.
REMOVE blister pack and CLIP the filter to the
C-Clips so the flow symbol on filter is toward the
machine’s inlet.
CONNECT the machine flex line to the outlet of
the filter.
OPEN the water supply valve and allow water to
run through the filter for 3-5 minutes to flush out
the disinfectant and purge air out of the system.
USE residual test strips on running water to
ensure residual disinfectant levels do not exceed
4ppm.
CHECK for any leaks and tighten any fittings if
necessary.
If included, PRESS ON the DSU Cover onto the
filter.
*It is recommended to install the DSU Cover to
shield the filter from the external environment.
MACHINE
FLOW
CONNECT
DSU
COVER
CONNECT
MACHINE
FLOW