
NEOGEN.com
10
Sample Procedure
01. To power on the instrument, press and hold the center button on your AccuPoint
®
Advanced NG. Enter user name
and pin if required.
02. Select standard test (when using a site plan) or quick test for general sampling. Under standard testing, select the
correct site to be tested.
03. Sample your site by drawing a 4 x 4 in (10cm x 10cm) square. With a vertical and horizontal back and forth motion
fill in the square. Consistency is critical.
04. Hold the sampler vertically and insert into unibody, and fully depress to activate.
Note:
The sampler must be
held vertically.
05. Swirl gently for two seconds.
Note:
Do not shake.
Preparing the Instrument for Use