Appendix
A-3
A4.1 Column Headings
NPB-190
Version 1.0.0
CRC XXXX
SpO2 Limit: 30-100%
PR Limit: 100-180 bpm
Time Tag
%SpO2
PR (bpm)
PA
Status
To explain the printout, it will be necessary to break it down to its key
components. The first two lines of the chart are the column headings shown
above. Every 25th line will be a column heading. A column heading is also
printed whenever a value of the column heading is changed. There are three
column heading lines shown in Figure A-2. Using the top row as the starting
point, there are 25 lines before the second column heading is printed. The third
column heading was printed because the SpO
2
limits changed from 30-100% to
80-100%.
Printout Source
NPB-190
Version 1.0.0
CRC XXXX
SpO2 Limit: 30-100%
PR Limit: 100-180 bpm
Time Tag
%SpO2
PR (bpm)
PA
Status
Data in the highlighted box above represents the source of the printout, in this
case the NPB-190.
Software Revision Level
NPB-190
Version 1.0.0
CRC XXXX
SpO2 Limit: 30-100%
PR Limit: 100-180 bpm
Time Tag
%SpO2
PR (bpm)
PA
Status
The next data field tells the user the software level, (Version 1.0.0) and a
software verification number (CRC XXXX). Neither of these numbers should
change during normal operation. The numbers will change if the monitor is
serviced and receives a software upgrade.
Alarm Limits
NPB-190
Version 1.0.0
CRC XXXX
SpO2 Limit: 30-100%
PR Limit: 100-180 bpm
Time Tag
%SpO2
PR (bpm)
PA
Status
The last data field in the top line indicates the high and low alarm limits for
%SpO
2
and for the pulse rate (PR). In the example above, the low alarm limit for
SpO
2
is 30% and the high alarm limit is 100%. Pulse rate alarm limits are: low
100 bpm, and high 180 bpm.
Column Headings
NPB-190
Version 1.0.0
CRC XXXX
SpO2 Limit: 30-100%
PR Limit: 100-180 bpm
Time Tag
%SpO2
PR (bpm)
PA
Status
Actual column headings are in the second row of the column heading line.
Patient data presented in the chart is, from left to right: the time that the chart
was printed, the current %SpO
2
value being measured, current pulse rate in beats
per minute (bpm), current pulse amplitude (PA), and the operating status of the
NPB-190.