Pericalm™ Instruction Manual
Disposal
Applications
Information on disposal for users of Waste Electrical &
Electronic Equipment (WEEE) for private households:
Electrical and electronic products including batteries should not
be mixed with general household waste. For proper treatment,
recovery and recycling, please take these products to designated
collection points, where they will be accepted on a free basis.
•
Promotes continence
•
Increases muscle strength
•
Maintains or improves range of movement
•
Increases and improves the blood supply to the muscle
•
Reduces pain
12
Pericalm™ Instruction Manual
Specifications
13
1.
Dual channel: individually isolated circuits.
2.
Amplitude: 0 - 90 mA; indication only; actual mA will tend
to be less than indicated due to electrode impedance.
3.
Type: Constant Current.
4.
Waveform: Asymmetrical, rectangular bi-phasic with zero
DC current.
5.
Selectable pulse width: 50 μS – 450 μS (2% accuracy)
6.
Pulse Rate selection: in the continuous mode 2 – 100 Hz
(2% accuracy)
7.
Time duration of the treatment: 5 - 60 minutes.
(Custom programmes only)
8.
Low Battery Indicator: If the battery goes below 6.9 volts +/- 0.2
volts the battery symbol will flash on/off once every second.
9.
Open Electrode Detect: If an open circuit is detected at the
output of channel A or B by: a probe not being attached,
or with a probe attached but not in position, when the unit
intensity is ramped up to 5mA the output current will be
reset at zero.
10.
Ramp up time 0.3 - 9.9 seconds.
11.
If the battery voltage is below 6.6 (+/- 0.2) volts the unit will
not turn on.
12.
Physical dimensions: 80 x 67 x 45 mm.
13.
Weight: 90 grams without battery, 125 grams with battery.
14.
Environmental conditions for storage & transport: -10 to +50
degrees Centigrade 0-90% Humidity.
Содержание PERICALM