English-26
Figure S.7
Figure S.8
Figure S.6
Figure S.9
7. The procedure will ask for the USB sensor to be placed on the center of the display panel (
Figure S.7
). Tilt the display
panel approximately 5˚ backward and place the USB sensor in the center of the display panel (
Figure S.1
).
NOTE:
Place the USB sensor flat against the LCD to avoid external light contamination.
DO NOT
press the calibrator
against the display panel. Press “SELECT”.
8. Set the target LUMINANCE (cd/m
2
) (
Figure S.8
).
9. Press “SELECT” to start the calibration. Stand-alone calibration may take several minutes depending on user settings.
10. After the CALIBRATION SUCCEEDED message appears (
Figure S.9
), press SELECT.
11. To end the calibration mode, press “EXIT”.