in- and out passage
function enabled
in- and out passage
function disabled
In Passage Delay
If the
in passage delay
is enabled, DOOR waits five seconds before
sending an alarm when the door is opened. During this time, the LED
on DOOR blinks with red light. If the power switch is turned off when
the LED is blinking, no alarm will be sent.
This way, care personnel can enter the room without an alarm being
sent. When care personnel leaves the room the power switch must be
turned on again.
Out Passage Delay
If the
out passage delay
is enabled, DOOR does not react to the door
being opened during the first five seconds after the power switch
on DOOR is turned on. This allows care personnel to leave the room
without an alarm being sent.
Configuration
The in- and out passage delay can be enabled by fitting the jumper on
the circuit board inside DOOR over both pins.
To change other settings, a computer with the DOOR configuration
software must be used.
Note:
The
in passage
and
out passage delay
do not apply to the external sensor.