P - 17
Susceptibility check
IEC 60601 test
conditions
Compliance
level
Electromagnetic environment-guidelines
Conducted RF IEC
61000-4-6
3 Vrms
150 kHz ~ 80 MHz
3 A/m
Portable and mobile radios should not be located near any part of the product,
including cables, that is closer than the recommended separation distance
calculated from the equation for the transmit frequency. (Recommended
separation distance : Not applicable)
80 MHz ~ 800 MHz
800 MHz ~ 2.5 GHz
P is the rated maximum output power (W) specified by the transmitter
manufacturer and d is the recommended separation distance (m). The
electromagnetic field strength c of the fixed RF transmitter, determined by the
electromagnetic site survey, must be less than the compliance level for each
frequency range. Disturbance may occur in the vicinity of medical devices
marked with the symbol below.
Radiated RF IEC
61000-4-3
3 V/m
80 MHz ~ 2.5 GHz
3 V/M
※ Note1
The higher frequency range applies in the 80 MHz and 800 MHz ranges.
※ Note2
These guidelines do not apply in all situations. Electromagnetic transmission is reflected and absorbed by buildings, objects, and
people