Bondascope 350 Operator's Manual
19
10.2 PITCH-CATCH Sample Setup Procedure
This is to provide a brief description for setting up and using a Pitch/Catch probes on the Bondascope 350/350.
$
Select the proper Pitch/Catch probe for the application
$
Select the BL8-BL8 or BF11-L8 (350) Cable for use. Attach probe to Cable.
$
Plug Cable into appropriate connector socket located on the top panel of gage.
$
Turn gage on.
If AUTO-PROBE is on the screen will show that the probe part number (i.e. PCM32, etc.) has automatically been
selected. If the probe is not automatically selected, a probe list will appear on screen to perform the following step(s):
$
Use the up/down arrow keys or the encoder knob to highlight “AUTO-PROBE” on the display screen. Press
“ENTER” on the keypad to turn on function. The screen display will show the probe part number of the probe
selected.
OR
$
Use the up/down arrow keys or the encoder knob to highlight the probe part number from the list of probes.
Press “ENTER” on the keypad to turn on function. The screen display will show the probe part number of the
probe selected. (Note: You will not be in the AUTO-PROBE Mode)
$
Use the up/down arrow keys or the encoder knob to highlight “Defaults”’ for a first time setup or “LAST” for a
previous non saved setup. Press “ENTER” (Stored setups can accessed in this matter.)
CALIBRATION
To set probe Pitch/Catch frequency :
$
Press “SWEEP” on the front panel.
$
Press “NULL” on the front panel.
$
Place probe on “Good” area of cal standard. Press “ENTER” on the front panel. Next, place probe on defect
area of cal standard. Press “ENTER”.
Press “RF/DOTS” on front panel to display RF. Adjust Gain, Range ,Delay , Cycles to get best defect response.
10.3 Care and use of Pitch-Catch Probes
Pitch-Catch probes are sensitive measurement devices. Use care when handling them The tips are sensitive and if
dropped, will likely, significantly damage the probe beyond use. Also, at the very least the tip may be broken within the
Transmit or Receive tower if care is not used. This is not a user serviceable part and is not covered under warranty.
The probe can be sent back to the factory for repair at prevailing rates.
To replace the tips when worn, simply pull them from their housings and replace with new tips. It may be helpful to
moisten the tip prior to insertion to settle into the internal suspension rings. DO NOT FORCE the tip in or element
damage WILL occur.
Probes offer several modes of use. There are 6 threaded holes on the bottom surface. To use the probe more within a
spring loaded, constant pressure condition, we install 2 guide feet just behind the probe tips and a single foot at the
rear (cable) end of the probe. This provides a relatively stable scanning position. On the other hand, if you prefer to
scan with more control on probe pressure move the guide feet into desired position. For instance, a single foot in the
rear will give a standard Tri-Point configuration providing a stable scanning condition.