AVENGER II User Manual
8 |
P a g e
NDT Systems, Inc.
5542
Buckingham Drive
,
Huntington Beach, CA 92649
,
PH: 714-893-2438
Menu 2 –
MONITOR “GATE a” functions.
GATE a
: Gate can be changed to PLOGIC, NLOGIC, CRV M, CRV-1, CRV-2, CRV+1,
CRV+2, EXPAND or OFF
PLOGIC
: When PLOGIC is selected, the MONITOR triggers if an echo is present
within the gate and its height is equal to or crosses the set Threshold Level.
NLOGIC
: When NLOGIC is selected, the MONITOR triggers if there is no echo
within the gate or the height is lower than the Threshold Level.
CRV M
: When DAC is turned ON and CRV M is selected the monitor triggers
if any Echo signal equals or crosses main DAC Curve
CRV-1
: If DAC is turned ON and if any echo signal cross DAC-1 dB curve then
the monitor triggers.
CRV-2
: If DAC is turned ON and if any echo signal cross DAC-2 dB curve then
the monitor triggers.
CRV+1
: If DAC is turned ON and if any echo signal cross DAC+1 dB
curve then the monitor
triggers.
CRV+2
: If DAC is turned ON and if any echo signal cross DAC+2 dB curve then the monitor
triggers.
EXPAND
: When EXPAND is selected the scale expansion activates and GATEa width expands to Full-
screen width for a magnified view.
STARTa
: GATEa starting position can be set using START. By pressing LEFT/RIGHT Keys Gate start can
be set in term of mm or in.
WIDTHa
: Gate width can be set using WIDTH, set in terms of mm or in.
LEVELa
: Gate Threshold Level, set in terms of the percentage of the Full Screen Height (FSH).
Example:
The MONITOR to be triggered when any echo crosses the 50% FSH between 35mm
and 80mm
.
•
In the Gate setting MENU 2, select PLOGIC using LEFT/RIGHT keys.
•
Set Gate START to 35mm using LEFT/RIGHT keys.
•
Set Gate END to 80mm using LEFT/RIGHT keys in WIDTH.
•
Set Gate LEVEL to 50% using LEFT/RIGHT keys.
GATEa
OFF
STARTa
30.0 mm
WIDTHa
40.0MM
LEVELa
30 %
GAIN 1
25.0 dB