Menu 13 –
ENCODER Inputs and Outputs.
ENCODER
: When Encoder is set to ON, Encoder value will be displayed.
CAL FAC
: Calibration Factor value is needed to calibrate the encoder.
Encoder Calibration:
•
Turn ENCODER to ON by holding CENTER key, current position value will set
to “0”.
•
Move encoder for known distance, about 100 to 500 mm.
•
Set displayed position value to actual travel distance value by adjusting “CAL
FAC”. Check again by moving encoder for known distance so it will display
same value.
ENCODER
ON
CAL FAC
1000
BSCAN D
ON
GAIN 1
25.0 dB