ZERO & VELOCITY
The next two functions, VEL and ZERO are adjustments necessary to "calibrate" the gage to the
accurately measure the specific material under test.
In our example, using a 0.500 inch thickness step, we have yet to perform any operations to
"calibrate" the Thickness readout. Most probably the reading is close to 0.500 inch, since the
FACTORY setup defaults are deliberately designed to produce nearly correct calibration for a
contact transducer being used on steel having longitudinal sound velocity of 0.2330
inches/microsecond.
21.
As a first step in becoming familiar with the VEL and ZERO functions, lets complete the
calibration for thickness measurement using the step block in our example.
VEL - Set the cursor at VEL and adjust it to 2330 In/us.
ZERO - Using the Left/Right arrow, reposition it at ZERO, probe zero control position. Using the
Up/Dn arrow keys adjust the thickness reading to 0.500". When 0.500 inch [or whatever the
known actual thickness of the step] is achieved in Thickness readout, the upper range
calibration is complete. Note that zero may change through up to 5 or 10 counts (on the display)
before the indicated thickness changes. Choosing zero roughly at the mid-point of the range
needed to move from 0.499 inch to 0.501 inch. This procedure can improve thickness gaging
precision.
22.
Remove the transducer from the 0.500 inch step and couple it to the other steps on the block. If
the test block has been accurately measured and the material is steel with a VELOCITY (VEL)
of 0.2330 in/us, the other step measurements should be within ±0.001 inch of the actual
measured values and reproducible.
21.
If some material other than steel is used, a different procedure will be required. For a
transducer of similar type as used in these procedures, reset the ZERO to approximately
that of the default (130). Depress the up arrow key to position the cursor at VEL (velocity).
With the transducer coupled to the thickest step, use the Up & Down arrow keys to
increase/Decrease the velocity until the known thickness of the material is displayed on
the thickness readout.
NOTE : to measure to thinnest step on the block it may be necessary to adjust IP-
BLK as per step 15 above