14 | Dome GX2MP Plus
OSD Main Menu
Only menu controls related to operation the Dome GX2MP Plus display using DVI-D
input are presented here. NOTE: Certain commands are inactive when DICOM Gamma
on the Color menu is set to BLS (backlight stabilization). All units ship with BLS enabled.
Main Menu
Submenu
Description
Brightness
Set brightness value from 0 to
100.
Contrast
Set contrast value from 0 to 100.
Sharpness
Set sharpness value from 0 to 10.
Color Temp
Set to Preset or User for screen
color adjustment.
Preset: Set to Warm (reddish),
Medium (sRGB standard color
specification), or Cool (bluish).
User: Set Red, Green, or Blue
color level from 0 to 100.
Gamma
Set black-white level of response
output to Gamma 0 (black),
Gamma 1, or Gamma 2 (white).
DICOM Gamma
Set gamma values for medical
characteristics and luminance:
On, Off, or BLS.
NOTE: Select On to maintain
luminance at the factory default
value. Select BLS for an internal
backlight sensor to maintain
luminance at 185 cd/m
2
. All units
ship with BLS enabled.