2030434-006 Rev. E
EasyOne-CS Field Service Manual
Page 7
Figure 8: Top view of EasyOne-CS PCB (without EMI shielding)
6
Specifications
6.1
EasyOne-CS Model 2001 Spirometer
Size:
83 x 158 x 43 mm (3.3 x 6.2 x 1.7 inches)
Weight:
242 grams (8,6 ounces), with batteries
Measuring accuracy
Volume: ±2% or 0.050 l
Flow:
±2% or 0.020 l/s, (except PEF)
PEF:
±5% or 0.200 l/s
MVV:
±5% or 5 l/min.
Measuring range:
Volume: ±12 l
Flow:
±16 l/s
Resistance:
below 0.3 cm H
2
O/L/s
Display:
64 x 160 graphic display
Data entry:
14-key keyboard
Data memory:
For up to 700 tests, stored in EasyOne CS
Test modes:
FVC, FVL, Slow VC, MVV, Pre/Post
Parameters:
FVC, MVV, FEV6, FEV1, FEV1/FVC, FEV1/VC , FEV1/FEV6, MEF25 (FEF75),
MEF50 (FEF50), MEF75 (FEF25), MEF25%-75%, PEF, FET, FIVC, PIF, IVC, ERV,
IRV, pre-post % variation, Lung Age
Respiratory tube:
Disposable spirette respiratory tube
Measurement principle:
Ultrasound transit-time measurement
Adult predicted:
NHANES-III, Knudson_83, Knudson_76, Crapo, Morris, ERS (ECCS/EGKS),
Forche (Austria), Sapaldia (Switzerland), Roca (Spain), Berglund, Gulsvik,
Hedenström, Gore, Cherniak (only for MVV)
Pediatric predicted:
Dockery (Harvard), Hsu, Zapletal, Polgar, Hibbert
Power supply
Power consumption:
2 alkaline batteries, Type AA, 1.5V
Typical 0.6 W
Report:
A4
Storage:
Temperature: -20 to 50 °C
Relative humidity: 5% to95%
Ambient pressure: 500 to 1060 hPa
Содержание EasyOne-CS
Страница 2: ......
Страница 29: ...2030434 006 Rev E EasyOne CS Field Service Manual Page 27 ...