18
N D B - D O C - 3 - U s e r M a n u a l
8. Battery replacement
The low battery indicator will turn on (red) when the battery
level is critically low. To replace the batteries, simply remove
the rubber protective case, slide the battery cover and carefully
remove the depleted batteries. Install a fresh set of four batter
-
ies. Refer to technical specifications for more information.
Make sure to respect the battery polarity. Reversing battery polarity can cause
leakage, overheating, create fire or explosion.
Do not attempt to charge alkaline batteries. They may leak, overheat, create
fire or explode.
9.
Technical Specifications
Technical Specifications
IP rating / IK rating
IP40 / IK06
Operating temperature
-20°C to 60°C
Dimensions
17.7 x 8.9 x 4.5 cm (7 x 3.5 x 1.8’’)
Weight
0.47 kg (1.03 lbs)
Power
4x AA 1.5V alkaline battery
Autonomy
8h of measurement
Measurement
Low voltage AC and DC
Nb of Channels
3
Detection
Short, Good, Open
Transformer power rating
1 kVA to 3000 kVA
Transformer nominal voltage rating
1000 V to 50000 V
Verification Features
Cable auto-verification
Transformer primary configurations
Single, Delta, Yn
Result indication
Audible, Visual (bright LEDs)
Note: Technical specifications can change without notice.