XDP-II User’s Guide
NDB Technologie Inc.
02/04/08
17
1. Plug the filter integrator (XDP-205) direcly to the input of XDP-II and slide the
switch to the 0 dB position.
2. Connect the calibrator and the XDP-II. The device under test as well as the
high voltage transformer source must be properly connected for valid
calibration.
3. Make sure that the XDP-202 is connected to the coupler and select
CALIBRATION PICO
menu and then 200pC.
Select 200pC