NAVAIR 17-20AQ-347
19
CALIBRATION CHECKLIST
TEST INST (S) Wavetek 1271 Digital Multimeter with or without OPT12 or 20
PROC. NO. NA 17-20AQ-347
MFR
MODEL
SER. NO.
PROCEDURE
MEASURED
VALUES
OUT
STEP
FUNCTION TESTED
NOMINAL
FIRST RUN
SECOND RUN
OF
CALIBRATION TOLERANCES
NO.
TOL
(1)
(2)
(3) (4) (5)
(6)
(7)
4.1
Direct
Voltage
Tests
4.1.1
Zero
Tests
4.1.1.2
TI Range (Zero Input)
(V)
(V)
”
100
mV
0.0000
-3
-0.0090
-3
to +0.0090
-3
”
1
V
0.000000
-0.000009
to
+0.000009
”
10
V
0.00000
-0.00009
to
+0.00009
”
100
V
0.0000
-0.0009
to
+0.0009
”
1
kV
0.000
-0.009
to
+0.009
4.1.4
Range
Divider
(V)
(V)
”
100
mV
0.0005000
10.0000
-3
9.9906
-3
to 10.0094
-3
”
” 0.0050000
100.0000
-3
99.9869
-3
to 100.0131
-3
”
1
V
0.0050000
0.100000
0.099989
to
0.100011
”
”
0.0500000
1.000000
0.999967
to
1.000033
”
10
V
0.0250000
0.50000
0.49990
to
0.50010
”
”
0.0375000
0.75000
0.74990
to
0.75010
”
”
0.0700000
1.40000
1.39988
to
1.40012
”
”
0.1300000
2.60000
2.59986
to
2.60014
”
”
0.2550000
5.10000
5.09981
to
5.10019
”
”
0.4000000
8.00000
7.99976
to
8.00024
”
”
0.5000000
10.00000
9.99972
to
10.00028
”
”
0.7000000
14.00000
13.99964
to
14.00036
”
”
0.9500000
19.00000
18.99955
to
19.00045
”
”
0.9500000
-19.00000
-18.99955
to
-19.00045
4.1.11.3
100
V
100.000
99.9961
to
100.0039
”
1000
V
1000.00
999.961
to
1000.039
*Select 10 Hz ACV filter.
Page 1 of 3