(18 blank) 17
4.4.8 Repeat the sections of the TI calibration that had previously failed. If these sections are completed satisfactorily, then
proceed to Section 4.5. Otherwise, consult the manufacturer’s manual to troubleshoot the problems encountered.
4.5 SECURING THE TI
4.5.1 Any steps that produced an out-of-tolerance reading must be repeated at this point. Use the TI SECOND RUN
checklist column to repeat the necessary steps and verify that these steps are now within the given tolerance limits.
4.5.2 Affix a SPECIAL CALIBRATION label to the TI indicating the following: “TI calibrated to Fluke 8505/8506 12
months specs (MSD assigned tolerance).” The date of calibration may by added to the TI memory, if desired, by following the
instructions in the manufacturer’s manual.
4.5.3 Unless other measurements are to be performed, set all power controls to off or standby, and disconnect the TI from the
setup.