Motorisation Technicals Characteristics
STARTING FROM SERIAL NUMBER W4ECP-1153
A. LCM DISPLAY
0. Information display
- To view the information registered
in the control box, you must press the
up and down buttons of the box for 5
seconds.
1. Operating factor
- Definition: It accounts for the actua-
tor (up / down) any breach of the duty
cycle = non-respect of the actuator rest.
2. Overloading of the actuator / a
- It counts the number of times an
attempt to lift a person weighing
beyond the maximum weight limit is
performed. When it occurs, the actua-
tor will stop and an anvil will appear, it
will be impossible to continue.
3. Overloading of the actuator / b
- It counts the number of times an
attempt to spread or close the opening
legs while they are stuck is performed.
When it occurs, the actuator will stop
and an anvil will appear, it will be
impossible to continue.
4. Incomplete load
- It records all incomplete loads below
24,6V. This corresponds to the number
of charges damaging the battery.
5. Time of use
- Definition: It displays the cumulative
operating time of the device. Starting
from 600,000 seconds of use, a mainte-
nance key is displayed; it is then essen-
tial to carry out a complete control of
the device.
22
User Manual - Stand-Up Lift: WAYUP 4 with electrical-opening legs
Manufactured by NAUSICAA Médical S.A.S. / Approved by Ghizlane Labrosse (Biomedical Engineer)