background image

User Manual 

 

Xltek EMU40EX Battery Pack

 

 

INTENDED USE 

The EMU40EX Battery Pack is intended for use with EMU40EX Breakout Box only, which is used in 
combination with NeuroWorks

®

/SleepWorks™ software to allow a qualified medical practitioner trained in 

electroencephalography or polysomnography to acquire, display, archive, review, and analyze 
physiological signals. It can be used with patients of all ages, but it is not designed for fetal use. 

INTENDED USERS AND PATIENT TARGET GROUP 

The intended users of the EMU40EX battery pack are trained medical professionals or someone 
instructed by a trained medical professional. At minimum, it is expected that the user has a basic 
understanding of clinical EEG and PSG, conducting an EEG or PSG study, proficiency with computers 
and modern operating conventions. The user is typically trained on the use of the 
NeuroWorks/SleepWorks software to operate the system or conduct a study acquisition using the 
EMU40EX amplifier. 

CLINICAL BENEFITS 

The EMU40EX batter pack extends the period that a patient may be in an ambulatory position or not 
physically connected to an amplifier base through a cable to facilitate the continuous recording of a 
patient’s physiological parameters. 

CONTRAINDICATIONS AND SIDE EFFECTS 

There are no known contraindications or side effects for procedures performed with the EMU40EX battery 
pack. 

CHARGING THE BATTERY PACK 

 

Connect the Battery Pack (Fig.1) to USB Power Adapter (Fig.2) or to a computer USB port using USB 
Power Cable (Fig.3). 

 

Check that “Charging” 

yellow light

 is on. If 

yellow light

 is off, it means, that charger does not detect 

sufficient power and will not charge properly. Check cable connections and try again. 

 

Leave the Battery connected to the power source until continuous 

yellow light

 

(“Charging”) is turned 

off and continuous

 green light

 

(“Battery”) is turned on. 

 

Unplug USB Power Cable (Fig.4) from the Battery Pack. All lights should be off. The Battery Pack is 
now ready to use. 

NOTE: Complete charging  cycle may take about 3½  hours  when  using USB  Power Adapter (Fig.2) and 

5~6 hours when using PC USB port. Charging the Battery Pack from USB port is not recommended 
due to longer charging time.  

Содержание Xltek EMU40EX

Страница 1: ...EMU40EX Battery Pack User Manual...

Страница 2: ...or in part except with the prior written permission of Natus Neurology Incorporated The copyright and the foregoing restrictions on the copyright use extend to all media in which this information is p...

Страница 3: ...D USERS AND PATIENT TARGET GROUP 7 CLINICAL BENEFITS 7 CONTRAINDICATIONS AND SIDE EFFECTS 7 CHARGING THE BATTERY PACK 7 USING THE EMU40EX HEADBOX WITH THE BATTERY PACK 8 USEFUL TIPS 8 MAINTENANCE 9 SP...

Страница 4: ...ording to IEC 529 The Battery Pack is not water proof Prevent detergent solution or cold sterilization agents from seeping into the electronics of the system Battery pack stored by plugging into the c...

Страница 5: ...to 50 75 before storage i e charge it for about 2 hours Device dropped or damaged in transit use could lead to loss of function or delayed diagnosis Inspect the device prior to each use and do not use...

Страница 6: ...Xltek EMU40EX Battery Pack User Manual 6 Fig 1 Battery Pack Fig 2 USB Power Adapter Fig 3 USB power Cable Fig 4 EMU40EX Power Cable...

Страница 7: ...The EMU40EX batter pack extends the period that a patient may be in an ambulatory position or not physically connected to an amplifier base through a cable to facilitate the continuous recording of a...

Страница 8: ...u do not intend to use the Battery Pack for a long time When the EMU40EX Power Cable is unplugged the Battery Pack automatically enters low power standby mode prolonging battery life All lithium batte...

Страница 9: ...acity 2500 2700mAh Energy capacity 9 8Watt hours Input charging voltage 5 12VDC Charging current limited to 500 1000mA 4 2V Charging time 3 hrs Charge retention storage 80 at Humidity 65 20RH 20 to 20...

Страница 10: ...ons state that electrical and electronic waste must be separately collected for the proper treatment and recovery to ensure that WEEE is reused or recycled safely In line with that commitment Natus ma...

Страница 11: ...State in which the user and or patient is established Refer to the Natus website for an electronic copy of this document INSTRUCTIONS TO ACCESS THE eIFU A copy of the Instructions for Use in PDF form...

Страница 12: ...ation to be supplied Manufacturer Indicates the medical device manufacturer ISO 15223 1 Symbol 5 1 2 Medical devices Symbols to be used with medical device labels labelling and information to be suppl...

Страница 13: ...1 Table D 2 2 Medical electrical equipment Part 1 General requirements for basic safety and Essential performance General warning sign Indicates a hazard of potential personal injury to patient or op...

Страница 14: ...Website natus com 033980 Rev 1 03 2021 A Total Service Solution Natus Systems are backed by a comprehensive extendable warranty Our support team is available around the clock Our technical staff prov...

Отзывы: