User & Service Manual
Xltek EMU 128 Headbox
8
Warnings and Cautions
General Warnings
Proper use of this device depends on careful reading of all instructions and
labels that come with or on the system. Inaccurate Measures may be caused by
incorrect application or use.
The NeuroWorks system is classified as an IP0 – ordinary degree of protection
against ingress of water according to IEC 529.
Inaccurate measurements may be caused by incorrect application or use.
Only qualified personnel should operate this equipment.
The computer used with the NeuroWorks system must either be approved by
Xltek and supplied as part of an IEC 601 approved system or it must be
approved to IEC 950 or similar and kept outside of the patient environment.
To ensure the validity of signals, do not operate the device near any sources of
electromagnetic interference.
Turn off the system power before cleaning. Prevent detergent solution or cold
sterilization agents from seeping into the electronics of the system. Be careful
around all connectors and edges. Do not use abrasive agents.
Explosion Hazard: This system is not AP or APG rated. DO NOT USE this
system in the presence of a flammable anesthetic mixture with air, oxygen or
nitrous oxide.
The accessories of this device may include several kinds of disposable, sterile
needle electrodes. These needles are labeled as STERILE and the method of
sterilization is documented on the packaging. These electrodes should not be
used if the sterile packaging has been tampered with.
The sale, distribution or use of this device is restricted to, by, or on order of a
physician.
Electrical Warnings and Cautions
The system is intended for connection to a properly grounded electrical outlet
only
Periodically check the
system ground integrity
, the
system leakage current
, and
the
patient contact leakage current
. This should be performed at least ONCE
PER YEAR.
Do NOT turn on the system power until all cables have been connected, verified
and visually inspected for any damage.
Failure to inspect the cables may result in
electrocution.
Do NOT connect or disconnect the patient headbox with the system power on.
ELECTRICAL SHOCK HAZARD: Do NOT connect electrode inputs to earth
ground. The patient headbox contains warning symbols reminding you that the
connections are intended for isolated patient connections only. Connecting to an
earth ground might result in electrocution.
ELECTRICAL SHOCK HAZARD: Do NOT
service the system. Refer servicing to
qualified personnel only.