background image

5.11

How to test the directional microphone

Directionality measurements as described in the hearing instrument test standards cannot be performed with regular desk-
top test chambers such as AURICAL HIT. Such measurements require large anechoic chambers. Small test chambers always
exhibit acoustic reflections that obscure the true directional behavior of the hearing instrument.

However, in AURICAL HIT you can make a functional test of the directional microphone in a hearing instrument. In this
test, the signal is first presented to the front of the hearing instrument and then to the back of the hearing instrument.
This is done automatically when you start a directional test. The signal used for this test is a flat spectrum Broad Band
Noise, band-pass filtered between 750 Hz and 5 kHz, and presented at 70 dB SPL.

The procedure

1. Position the hearing instrument as described in

Traditional BTE hearing instruments

12

,

Thin-tube hearing instru-

ments

13

and

ITE hearing instruments

14

depending on the type of hearing instrument.

2. In the field

Directionality Adaptation

in the OTOsuite HIT module, you can define the duration of the signal present-

ation before the actual measurement is made. This value accommodates any adaptational behavior of the hearing
instrument. Adaptive directionality often takes 10 to 15 seconds or more before the directionality of the hearing
instrument is fully efficient.

3. You can either combine the

Directional Test

with your standard test sequence by checking the sequence checkbox,

or run it separately by clicking the

Start

button.

The result

The

Directional Test

result is shown as a 1/3 octave curve of the difference between the measurement with noise presen-

ted from the main loudspeaker and the measurement with noise presented from the rear loudspeaker. The numerical dir-
ectionality result shown in the

Measurements

table indicates the average front/back difference in the measured

frequency range.

The

Measurements

table also includes the adaptation interval used in seconds.

6

Maintenance and Calibration

Warning

Under no circumstances disassemble AURICAL HIT. Contact your supplier. Parts inside AURICAL HIT must

only be checked or serviced by authorized personnel.

Calibration

Calibration of the coupler microphone, and calibration of a new reference or coupler microphone must only be performed
by authorized personnel.

Maintenance

AURICAL HIT requires no preventive maintenance except for cleaning and regular calibration of the reference microphone.

Repair

For any type of repair, please contact your supplier.

16

Otometrics - AURICAL HIT

6 Maintenance and Calibration

Содержание Otometrics Aurical HIT

Страница 1: ...AURICAL HIT and the OTOsuite HIT Module User Guide Doc No 7 50 1230 EN 06 Part No 7 50 12300 EN...

Страница 2: ...etrics the Otometrics Icon AURICAL MADSEN HI PRO 2 Otoscan ICS and HORTMANN are registered trademarks of Natus Medical Denmark ApS in the U S A and or other countries Version release date 2019 09 18 2...

Страница 3: ...5 3 Installation 5 4 The test chamber 5 5 Testing hearing instruments 9 6 Maintenance and Calibration 16 7 Other references 17 8 Technical specifications 17 9 Definition of symbols 20 10 Warning notes...

Страница 4: ...ument dispensers and other health care pro fessionals in testing programmable hearing instruments Required qualifications It is assumed that the user has a basic knowledge of how to compare the result...

Страница 5: ...ecessary parts and accessories If your package is incomplete contact your local distributor 3 Installation Place AURICAL HIT on an absolutely stable surface In order to exclude ambient noise and to co...

Страница 6: ...e 8 C Cable groove 8 The AURICAL HIT handle 8 some models only 4 1 The coupler assembly The coupler assembly consists of the following parts A Coupler adapter B Coupler cavity C Coupler microphone 6 O...

Страница 7: ...n 2cc coupler and ear sim ulator The selected coupler type is saved with measurements for later reference Coupler microphone The coupler microphone is located in a coupler bottom piece which must be a...

Страница 8: ...mitters and body worn hearing instruments at a level where the microphone or microphones are approximately centered in relation to the loudspeaker 4 4 The AURICAL HIT handle Note This only applies to...

Страница 9: ...ng instruments 12 Thin tube hearing instruments 13 ITE hearing instruments 14 3 Testing When you have positioned the hearing instrument correctly you can test it using the OTOsuite HIT module as descr...

Страница 10: ...trument the only two important things to remember are Aligning directional microphones along the loudspeaker axis Positioning the reference microphone as close as possible to the front microphone of t...

Страница 11: ...HIT automatically detects the battery simulator Caution When you have connected the battery simulator make sure that it does not touch other metal parts as this may short cir cuit the system 5 4 Coupl...

Страница 12: ...E hearing instruments This procedure applies to any type of standard BTE hearing instruments with traditional earmolds Using the HA 2 adapter and BTE adapter tube 12 Otometrics AURICAL HIT 5 Testing h...

Страница 13: ...the Receiver In the Ear RIE Receiver In the Canal RIC and pre bent tubing Using the HA 1 ITE adapter Note If you place some acoustic putty on the receiver wire this will shift its resonance frequency...

Страница 14: ...ing instrument During telecoil testing AURICAL HIT automatically detects the orientation of the hearing instrument 2 Enable telecoil mode in the hearing instrument 3 Close the lid and start testing 5...

Страница 15: ...tically when you use OTOsuite with Noah together with the fit ting software for the hearing instrument 5 Position the hearing instrument so that it is ready for testing and switch it on 6 If you wish...

Страница 16: ...accommodates any adaptational behavior of the hearing instrument Adaptive directionality often takes 10 to 15 seconds or more before the directionality of the hearing instrument is fully efficient 3 Y...

Страница 17: ...liquids Do not allow moisture inside the unit Moisture inside the unit can damage the instru ment and it may result in a risk of electrical shock to the user or patient Adapters If needed remove any a...

Страница 18: ...ncy range coupler microphone equalized 125 to 200 Hz 3 dB 200 to 5000 Hz 1 dB 5000 to 10000 Hz 3 dB Battery simulator Voltage range 0 to 2 0 V Resolution voltage 0 02 V Accuracy voltage 0 05 V Output...

Страница 19: ...et Warm up time 15 min Standards AURICAL HIT CE marked according to the Low Voltage Directive 2014 35 EU Safety IEC 61010 1 2010 Test standards ANSI S3 22 2009 IEC 60118 7 2005 EMC IEC 61326 1 2013 Ac...

Страница 20: ...ark USB socket for connecting AURICAL HIT to a computer Socket for connecting external coupler microphone Electronic equipment covered by the Directive 2012 19 EU on waste electrical and electronic eq...

Страница 21: ...y used for purposes other than the applications defined in the Intended Use section of this manual 10 The device should be turned off before any connections are established To disconnect the device fr...

Страница 22: ...The manufacturer is to be considered responsible for effects on safety reliability and performance of the equipment only if All assembly operations extensions re adjustments modifications or repairs a...

Отзывы: