Madsen AccuScreen OAE & ABR Screener User Manual
60
8.2.2
From the Quality Tests menu
1.
Insert the probe tip without eartip in the test cavity on the device, or in an
external test cavity.
2.
Press
Probe (
)
on the
Quality Tests
screen.
3.
For more information, see
8.3
Electrode cable test
Before you start testing patients, you should test the electrode cable to ensure that it functions correctly.
1.
Clip the electrode clips on to the ABR Tester metal bars.
2.
Press
Electrode cable
(
) on the
Quality Tests
screen.
If the electrode cables function correctly, the message
Electrode cable OK
appears.
If an electrode cable does not function correctly, the message
Electrode cable
failed
appears.
3.
Press
OK (
)
to confirm the test result.