S88X
Dual Output Square Pulse Stimulator
7-41
r
ePetitive
P
ulses
(s1
or
s2)
1.
Select
S1
or
S2
FUNCTION
, and use the thumbwheel to select
S1
(
or
S2
)
REPEATING PULSES.
2.
Depress the
TIMING
RATE
button to access the rate timing. Use the left
and right arrow keys and the thumbwheel to select a rate between
0.01
and
1000
PPS (Pulses per Second).
3.
Select the
TIMING
DELAY
button and use the left and right arrow keys
and the thumbwheel to set the desired delay value observing the
50% duty cycle limitation.
4.
Select the
TIMING
DURATION
button and use the left and right arrow
keys and the thumbwheel to set the desired duration value observing
the 50% duty cycle limitation.
5.
Set the
OUTPUT AMPLITUDE
in volts using the thumbwheel and the left
and right arrow keys. Voltages between
0.01
and
24
can be selected.
6.
Depress the
OUTPUT ON/OFF
button.
7.
To synchronize other devices to the stimulus, connect either the
S1
(
or
S2
)
SYNC OUT
(no delay), or the
S1
DEL
(or
S2
DEL
) if delay is required, to
the other device synchronous input.
s
iNgle
t
raiNs
(s1
or
s2)
1.
Select
S1
or
S2
FUNCTION
, and use the thumbwheel to select
S1
(
or
S2
)
FUNCTION = SINGLE TRAIN (ST).
2.
Select
TRAIN DURATION
button and set the duration of the train using
the left and right arrow keys and the thumbwheel between
0.001
and
99.999
seconds.
3.
Depress the
TIMING
RATE
button to access the rate timing. Use the left
and right arrow keys and the thumbwheel to select a rate between
0.01
and
1000
PPS (Pulses per Second). This rate will become the
number of pulses within the selected
TIMING
DURATION
.
4.
Select the
TIMING
DELAY
and use the left and right arrow keys and
thumbwheel to set the desired delay observing the duty cycle limit.
This delay will be the time from the
S1 Delay Sync Out
to each stimulus
pulse in the train.
5.
Select the
TIMING
DURATION
and use the left and right arrow keys and
thumbwheel to set the pulse duration observing the duty cycle limit.
6.
Select the
OUTPUT AMPLITUDE
voltage and select the
OUTPUT ON/OFF
to
on
.
7.
Depress the
S1
(
or
S2
)
SINGLE
button to initiate a single train.
Содержание Grass S88X
Страница 2: ......
Страница 10: ...x Table of Contents...
Страница 12: ...1 12 Chapter 1 Specifications Glossary of Symbols Warnings Cautions...
Страница 24: ...4 24 Chapter 4 Description of Circuits...
Страница 54: ...8 54 Chapter 8 Artifacts Stimulus Isolation Constant Current...
Страница 64: ...B 64 Chapter B Appendix...