![natus Echo-Screen III Pro Скачать руководство пользователя страница 51](http://html1.mh-extra.com/html/natus/echo-screen-iii-pro/echo-screen-iii-pro_instructions-for-use-manual_3504910051.webp)
© 2020 Natus Medical Incorporated
47
026057 RevE
Appendix A: Electromagnetic Compatibility (EMC)
Echo-Screen III Pro Instructions for Use
NOTICE: Deviations from the collateral standard were performed per IEC 60601-2-40
NOTICE: All necessary instruction for maintaining compliance with regard to EMC can be found
in the general maintenance section in this instruction. No further steps are required.
To ensure compliance with the EMC requirements as specified in IEC 60601-1-2, it is
essential to use only the following accessories:
NOTICE: Conformance to the EMC requirements specified in IEC 60601-1-2 is ensured if the
cable types and cable lengths are as specified above:
5,240
5,100 –
5,800
WLAN
802.11 a/n
Pulse modulation
217 Hz
0,2
0,3
9
5,500
5,785
ITEM
MODEL
MANUFACTURER
Cable length
[meters]
Echo-Screen III OAE Probe Cable
012625
Natus Medical
1.2 M
Echo-Screen III ATA Cable
010999
Natus Medical
1.2 M
Echo-Screen III PCA Cable
009149
Natus Medical
1.2 M
Universal AC Adapter
013401
GlobTek
1.83 M
USB Cable, A to Micro B
011029
L-Com
2 M
Test
Frequency
(MHz)
Band
(MHz)
Service
Modulation
Maximum
Power (W)
Distance
(m)
Immunity
Test
Level
(V/m)
Содержание Echo-Screen III Pro