© 2020 Natus Medical Incorporated
31
026057 RevE
Testing Equipment
Echo-Screen III Pro Instructions for Use
3
On the Systems Test screen, touch
Touch Panel Test
.
4
Touch the center of each with your finger. When a touch point passes, the orange
+ changes to a green +.
5
At the end of the test, touch
Done
.
6
If any of the touch points did not pass, contact Natus Technical Service or your authorized
service representative.
Keyboard
Test the keyboard:
1
On the Home screen, touch
Equipment Check
.
2
On the Equipment Check screen, touch
Systems Test
.
3
On the Systems Test screen, touch
Keyboard Test
.
4
Press the key on the device keyboard that corresponds to the key highlighted in blue on the
test screen. When a key passes the test, it turns green, and the next key is highlighted in
blue. If a key does not turn green, the key failed the test. Press the barcode button to end
the test, and contact Natus Technical Service or your authorized service representative.
To end the test early, press the barcode button on the front of the device.
5
When the test is complete, touch
Done
.