Dantec Keypoint G4
28
Stimulators
Safety Information
Stimulators
WARNING
When operating the current stimulators be careful not to expose patients to high currents.
Before connecting or disconnecting the stimulation electrode, always “reset” the stimulator.
WARNING
Simultaneous connection of a patient to HF surgical equipment may result in burns at the
site of the electrical stimulation, or recording electrodes and possible damage to the electrical stimulator, or
the electrode input amplifiers. Operation in close proximity (e.g. 1m) to short wave, or microwave therapy
equipment may produce instability in the electrical stimulator output.
WARNING
Dangerous physiological effects! The current stimulator may give off dangerous
currents and voltage.
CAUTION
Avoid electrical stimulation for an extended period of time.
Stimulation Electrodes
WARNING
When using needle electrodes for recording or stimulation, either use pre- sterilized
single-use needle electrodes, or thoroughly sterilize multi-use needle electrodes.
NOTE
For cleaning multi-use surface electrodes, refer to their accompanying instructions for use.
CAUTION
Avoid accidental contact between connected, but unapplied electrodes and other
conductive parts - including those connected to protective earth.