Steam Pre-Vacuum Sterilization Parameters
Parameters
Option 1 (US)
Option 2 (EU)
Option 3 (EU)
Specifications
Temperature
132 ºC (270 ºF)
134 ºC to 137 ºC
(273 ºF to 279 ºF )
134 ºC (273 ºF)
Exposure Time
4 minutes
3 minutes
18 minutes
Dry Cycle Time
20 minutes
16 minutes
20 minutes
The Option 2 and Option 3 sterilization cycles are not considered by the United States Food and Drug
Administration (US FDA) to be standard sterilization cycles. Users should only use sterilizers and accessories (such
as sterilization wraps, sterilization pouches, chemical indicators, biological indicators, and sterilization
containers) that have been cleared by the US FDA for the selected sterilization cycle specifications (time and
temperature).
Following Sterilization
Allow cables to cool down to ambient temperature before starting the monitoring process. Before each use,
visually inspect the sterilized cables for any possible defects. Do not use defective cables.
Containment and Transportation of Sterilized Cables
It is recommended that the FLEX Extension Cable be sterilized as soon as possible after use.
Storage of
Sterilized Cables
Sterilized cables should be stored at room temperature in sterile pouches.
Содержание Camino ICP
Страница 1: ...1 Camino ICP Monitor Intracranial Pressure and Temperature Monitor User s Manual 025870 Rev 00 01 2019...
Страница 2: ...2...
Страница 12: ...12 Blank page...
Страница 16: ...16 Blank page...
Страница 26: ...26 About the Left Panel Theleftpanel doesnotcontain anyusableconnector ports or buttons...
Страница 56: ...56 Blank page...
Страница 74: ...74 Blank page...
Страница 92: ...92 Blank page...
Страница 100: ...100 Blank page...
Страница 110: ...110 Blank page...
Страница 142: ...142 Blank page...
Страница 146: ...146 Blank page...