4
Installation and Maintenance Manual
4. DIRECTIONS FOR PROPER PLACEMENT OF THE COMPONENTS:
o
Place the unit in an appropriate location, far from heat sources and with
adequate ventilation, on a
fl
at level surface. The surface must be able
to support the 100 lb weight of the unit.
o
Locate the unit within 10 feet of a water supply.
o
Locate the unit within 6 feet of a 115 VAC, 15 amp GFCI power supply.
o
Locate the Filter Set within 5 feet of the unit.
o
Maintain at least 6 inches of clearance on both sides and 2 inches in back for
proper
ventilation.
o
For Remote Chiller Model: provide ventilated cabinet doors and sides to
allow for fresh air into the evaporator coil side and exhaust air out side.
Without proper ventilation the unit will not perform properly and will shorten
the life of the Compressor.
o
Allow 14 inches minimum above the unit for removing the Top to perform
routine maintenance,
and to allow for checking of the water level in the
Reservoir during routine maintenance.
o
The Water Leak Protection Device must be connected directly to the Braided
Inlet Water Line at the customer supplied valve.
6”
6”
2”
REMOTE CHILLER MODEL
6”
6”
2”
COUNTER TOP MODEL
MODEL
REMOTE CHILLER
MODEL