background image

  7 

3.  ELECTRICAL MAINTENANCE  
 

A.  The extent of electrical maintenance, by other than a qualified technician, should be bulb replacement.  

BE CERTAIN TO TURN SWITCHES OFF AND REMOVE PLUG FROM POWER SOURCE OUTLET 
BEFORE CHANGING BULBS. 

 

B.  To replace bulb, gently lay microscope on side. Using a Phillips (cross-head) type screwdriver, remove 5 

each 3mm Phillips screws from bottom of base.  Remove base plate to expose bulb.  Using a tissue or 
cloth to gently grasp the halogen bulb, pull straight out of lamp socket.  Your microscope requires a 12 
volt, 20 watt halogen bi-pin bulb, available from the dealer from which your purchased your microscope.  
This is a common microscope bulb, Osram #64425 or equal.  Make certain that new bulb is clean.  
Grasping bulb gently with tissue or cloth, insert pins straight into lamp socket.  Replace base, and secure 
with 5 each 3mm Phillips screws. 

 

TROUBLESHOOTING 

 

PROBLEM 

REASON FOR PROBLEM 

SOLUTION 

Light fails to operate. 

Outlet inoperative. 

Have qualified service technician 
repair outlet. 

 

 

 

 

AC power cord not connected. 

Plug into outlet. 

 

 

 

 

12 volt DC output plug not 
plugged into DC power jack. 

Plug 12 volt DC output plug into 
DC power jack. 

 

 

 

 

Lamp burned out. 

Replace lamp. 

Image does not remain in focus 
 

Stage of microscope drops from 
its own weight. 

Adjust tension control. 
 

Image will not focus 

Rack stop not set at proper 
position. 

Adjust rack stop. 

 

 

 

 

Slide upside down. 

Place slide on stage with cover 
slip up. 

 

 

 

 

Slide cover slip too thick. 

Use 0.17mm thick cover slip  
     (No.1 cover slip) 

Poor resolution 
     (image not sharp) 

Objective lenses dirty. 

Clean objective lenses. 

 

 

 

 

Eyepiece lens dirty. 

Clean eyepiece lenses. 

 

 

 

 

Too much light. 

Adjust disc diaphragm. 

Spots in field of view. 

Eyepiece or condenser lens dirty. 

Clean lens. *** 

 

 

 

 

Specimen slide dirty. 

Clean slide. 

***Spots in field of view can also result from dirt on inside of eyepiece.  It is recommended that you have 
service technician clean inside of lens. 

 
OPTIONAL ACCESSORIES AND PARTS: 

 
#610-155 

WF10x eyepiece w/pointer 

#800-160 

Replacement bulb, 12v 20 watt halogen bi-pin, (Osram #64425 or equal) 

 
LIMITED LIFETIME WARRANTY 

 

Please see our website, 
 
 
 
 
 
 
 
 
 
 
 
 

(Revised 2/18/09) 

Содержание DC4-156-S

Страница 1: ...ered in separate instructions HOW TO USE YOUR MICROSCOPE SERIAL NUMBERS 1 Microscope serial number This number etched on back arm of microscope is the number under which your warranty is registered 2...

Страница 2: ...holder mechanical stage Fine focus knob USB output terminal 12 volt DC input Main Switch 12 volt DC output 12 volt DC switching power supply Detachable power supply cord 100v 240v 50H 60H input Rack...

Страница 3: ...ransported or returned to factory for any reason If it becomes necessary to ship the microscope for any reason repack it in the styrofoam container and then pack the styrofoam in another corrugated sh...

Страница 4: ...he software documentation in order to utilize some of the unique features provided by the digital camera and software 1 Illumination A Adjust intensity control on side of base to the minimum position...

Страница 5: ...another while requiring only a slight adjustment of the fine focus controls c When changing to the 40x and 100x objective lens care must be exercised when positioning these lenses into the optical pa...

Страница 6: ...for the presence of such contaminants If a microscope is returned to National for warranty repair and it is determined that such contaminants are the problem this is not covered under warranty and Na...

Страница 7: ...hnician repair outlet AC power cord not connected Plug into outlet 12 volt DC output plug not plugged into DC power jack Plug 12 volt DC output plug into DC power jack Lamp burned out Replace lamp Ima...

Отзывы: