OBJECTIVE

 

 

 

DIAPHRAGM OPENING

 

4x 

 

 

From fully closed to 1/8 open 

10x 

 

 

1/8 to 1/4 open 

40x 

 

 

1/4 to 1/2 open 

100x 

 

 

1/2 to 3/4 open 

 

F.  Usage of filters (one 32mm O.D. transparent blue filter and one frosted neutral filter has been provided 

with your microscope) 

 

a)  Placing a filter into the optical path will absorb some of the light from the illuminator base, limiting the 

light output.  Filters should only be used when they help enhance the image or to help reduce 
brightness of field. 

 

b)  To insert filter, grasp filter holder knob and swing out filter holder, insert filter into slot provided and 

swing filter into optical path. 

 

c)  When handling the filter do not touch lens surfaces as fingerprints and smudges will result in poor 

images. 

 

G.  Changing magnification. 

 
a)  Rotate revolving nosepiece to position 10x objective into optical path. 

 

b)  This microscope has been parfocalized, which allows changes from one objective to another while 

requiring only a slight adjustment of the fine focus controls. 

 

c)  When changing to the 40x and 100x objective lens, care must be exercised when positioning these 

lenses into the optical path, in order to prevent damaging the front lens element and specimen slide. 

 

d)  In order to obtain maximum resolution of the 100x oil immersion lens,  it is necessary to apply 

immersion oil between the coverglass of slide and front lens of the objective. 

 

(1)  Use of a very small amount of immersion oil is required. 

 

(2)  All air bubbles must be removed from between lens and slide by gently rotating nosepiece back 

and forth. 

 
(3)  When finished viewing, all parts that come in contact with oil must be cleaned.  Failure to do so 

could permanently damage the 100x oil objective lens.  Use of Windex to clean immersion off 
lens surfaces is recommended. 

 

Microscope Specification Chart 

Objective 

N.A. 

Color Code 

Ring 

Field of 

View 

Working 

Distance 

Magnification with 

WF10X eyepiece 

Din 4X 

0.10 

Red 

4.5mm 

26.4mm 

40X 

Din 10X 

0.25 

Yellow 

1.8mm 

5.5mm 

100X 

Din 40X  

0.65 

Blue 

0.45mm 

0.48mm 

400X 

Din 100X 

1.25 

White 

0.18mm 

0.06mm 

1000X 

 

Содержание DC4-156-S

Страница 1: ...ered in separate instructions HOW TO USE YOUR MICROSCOPE SERIAL NUMBERS 1 Microscope serial number This number etched on back arm of microscope is the number under which your warranty is registered 2...

Страница 2: ...holder mechanical stage Fine focus knob USB output terminal 12 volt DC input Main Switch 12 volt DC output 12 volt DC switching power supply Detachable power supply cord 100v 240v 50H 60H input Rack...

Страница 3: ...ransported or returned to factory for any reason If it becomes necessary to ship the microscope for any reason repack it in the styrofoam container and then pack the styrofoam in another corrugated sh...

Страница 4: ...he software documentation in order to utilize some of the unique features provided by the digital camera and software 1 Illumination A Adjust intensity control on side of base to the minimum position...

Страница 5: ...another while requiring only a slight adjustment of the fine focus controls c When changing to the 40x and 100x objective lens care must be exercised when positioning these lenses into the optical pa...

Страница 6: ...for the presence of such contaminants If a microscope is returned to National for warranty repair and it is determined that such contaminants are the problem this is not covered under warranty and Na...

Страница 7: ...hnician repair outlet AC power cord not connected Plug into outlet 12 volt DC output plug not plugged into DC power jack Plug 12 volt DC output plug into DC power jack Lamp burned out Replace lamp Ima...

Отзывы: