Using the Integrated Panel Mount Feature
The NI WSN-9791 integrated panel mount feature accepts number 8 or M4 pan head screws to secure
the device onto a flat surface. To install the NI WSN-9791 on a flat surface, complete the following
steps:
1.
Prepare two holes in your panel 4.448 in. (112.97 mm) on center to accept the number 8 or M4 pan
head screws or bolts.
2.
Install the front screw first in the mounting panel leaving .08 in. space under the head of the screw.
3.
Slide the NI WSN-9791 front mount under this screw.
4.
Fasten the rear screw into place through the rear enclosure mount of the NI WSN-9791.
Figure 6.
Using the Integrated Mount Feature
4.500 in.