NI WSN-3226 Calibration Procedure

4.

Configure the NI WSN-3226 according to Table 2.

5.

Click 

OK

.

6.

Right-click the NI WSN-3226 in the Project Explorer and select 

Deploy

.

7.

Set the calibrator to a Test Point value indicated in Table 3.

8.

Create a VI in LabVIEW to acquire a voltage reading from the AI 0 variable on the 
NI WSN-3226.

9.

Acquire and average 25 voltage readings with the NI WSN-3226.

10. Compare the average of the NI WSN-3226 readings with the 3-Year Limits in Table 3.
11. Repeat steps 7 through 10 for all test points in Table 3.

Note

The limits in Table 3 are derived using the values in Table 6.

If the Voltage Accuracy verification procedure determines that the NI WSN-3226 is outside of 
the limits, refer to 

Where to Go for Support

 for assistance in returning the device to NI.

Table 2.  

NI WSN-3226 Voltage Configuration

Channel 

Attribute

*

Channel 

Value

*

Sample 
Interval

Powerline 

Filtering

Filtering 

Strength

Measurement 

Type

Voltage

1

50/60 Hz

High Rejection

Set the channel attribute and value for each analog input channel.

Table 3.  

NI WSN-3226 Voltage Test Limits

Range (V)

Test Point

3-Year Limits

Positive 

Full Scale

Negative 

Full Scale

Value 

(V)

Lower Limit 

(V)

Upper Limit 

(V)

10.000

-10.000

9.95000

9.94332

9.95668

0.00000

-0.0027

0.0027

-9.95000

-9.95668

-9.94332

Содержание WSN-3226

Страница 1: ...Accuracy Under Calibration Conditions 6 Where to Go for Support 7 Software Calibrating the NI WSN 3226 requires the installation of LabVIEW 2010 or later and NI WSN 1 2 or later on the calibration sy...

Страница 2: ...evice are secure The NI WSN 3226 must be powered by either an external 9 V to 30 V power supply or four AA 1 5 V batteries alkaline only LabVIEW Help LabVIEW programming concepts and reference informa...

Страница 3: ...llowing performance verification procedures describe the sequence of operation and test points required to verify the NI WSN 3226 and assume that adequate traceable uncertainties are available for the...

Страница 4: ...Table 3 Note The limits in Table 3 are derived using the values in Table 6 If the Voltage Accuracy verification procedure determines that the NI WSN 3226 is outside of the limits refer to Where to Go...

Страница 5: ...orresponds to the appropriate RTD resistance range 8 Turn any connection compensation settings on the calibrator off Tip On the Fluke 5520A press the COMP softkey to turn compensation off 9 Create a V...

Страница 6: ...racy Under Calibration Conditions The values in the following tables are based on calibrated scaling coefficients which are stored in the onboard EEPROM The following accuracy table is valid for calib...

Страница 7: ...Specifications online at ni com manuals Where to Go for Support The National Instruments website is your complete resource for technical support At ni com support you have access to everything from t...

Отзывы: