NI WSN-3226 Calibration Procedure

Test Equipment

Table 1 lists the equipment recommended for the performance verification procedures. If the 
recommended equipment is not available, select a substitute using the requirements listed in 
Table 1.

Test Conditions

Follow these guidelines to optimize the equipment and the environment during calibration:

For  the 

Resistance Accuracy Verification

 procedure, keep connections to the 

NI WSN-3226 under 5 inches and within 1/8 inch in length of each other.

Verify that all connections to the device are secure.

The NI WSN-3226 must be powered by either an external 9 V to 30 V power supply or 
four AA (1.5 V) batteries (alkaline only).

LabVIEW Help

LabVIEW programming concepts and reference information about NI-WSN VIs 
and functions

Table 1.  

Recommended Equipment

Equipment

Recommended 

Model

Where 

Used

Requirements

Calibrator

Fluke 5520A

Voltage

Use a high-precision voltage 
source with an accuracy of at least 
50 ppm and an output impedance 
of 

 1 

Resistance

Use a high-precision resistance 
calibrator with an accuracy of at 
least 90 ppm

WSN Ethernet 
gateway and 
power supply

NI WSN-9791 with 
9 V to 30 V power 
supply

All

NI 9792 with 9 V to 
35 V power supply

NI 9795 with 
CompactRIO 
system

Screw 
terminal 
wiring

All

14 to 24 AWG wire

Содержание WSN-3226

Страница 1: ...Accuracy Under Calibration Conditions 6 Where to Go for Support 7 Software Calibrating the NI WSN 3226 requires the installation of LabVIEW 2010 or later and NI WSN 1 2 or later on the calibration sy...

Страница 2: ...evice are secure The NI WSN 3226 must be powered by either an external 9 V to 30 V power supply or four AA 1 5 V batteries alkaline only LabVIEW Help LabVIEW programming concepts and reference informa...

Страница 3: ...llowing performance verification procedures describe the sequence of operation and test points required to verify the NI WSN 3226 and assume that adequate traceable uncertainties are available for the...

Страница 4: ...Table 3 Note The limits in Table 3 are derived using the values in Table 6 If the Voltage Accuracy verification procedure determines that the NI WSN 3226 is outside of the limits refer to Where to Go...

Страница 5: ...orresponds to the appropriate RTD resistance range 8 Turn any connection compensation settings on the calibrator off Tip On the Fluke 5520A press the COMP softkey to turn compensation off 9 Create a V...

Страница 6: ...racy Under Calibration Conditions The values in the following tables are based on calibrated scaling coefficients which are stored in the onboard EEPROM The following accuracy table is valid for calib...

Страница 7: ...Specifications online at ni com manuals Where to Go for Support The National Instruments website is your complete resource for technical support At ni com support you have access to everything from t...

Отзывы: