TB-9212 Calibration Procedure

|

© National Instruments

|

5

Test Conditions

The following setup and environmental conditions are required to ensure the TB-9212 meets 
published specifications.

Keep connections to the TB-9212 as short as possible. Long cables and wires act as 
antennas, picking up extra noise that can affect measurements.

Verify that all connections to the TB-9212 are secure.

Use thermocouple wire for all cable connections to the TB-9212. 

Maintain an ambient temperature of 23 °C ± 5 °C. The TB-9212 temperature will be greater 
than the ambient temperature.

Keep relative humidity below 80%.

Allow a warm-up time of at least 15 minutes to ensure that the NI 9212 measurement 
circuitry is at a stable operating temperature.

Initial Setup

Refer to the

 NI cDAQ-9174/9178 Quick Start 

for information about how to install the software 

and hardware and how to configure the device in Measurement & Automation Explorer (MAX).

Complete the following steps to set up the TB-9212.
1.

Install NI-DAQmx.

Note

Ensure that you install NI-DAQmx 15.1 or later. Calibrating the TB-9212 

using NI-DAQmx 14.5 or earlier may result in inaccurate readings.

2.

Make sure the cDAQ-9178 power source is not connected.

3.

Connect the cDAQ-9178 to the system safety ground.
a.

Attach a ring lug to a 14 AWG (1.6 mm) wire.

b.

Connect the ring lug to the ground terminal on the side of the cDAQ-9178 using the 
ground screw.

c.

Attach the other end of the wire to the system safety ground.

4.

Install the NI 9212 in slot 8 of the cDAQ-9178 chassis. Leave slots 1 through 7 of the 
cDAQ-9178 chassis empty.

5.

Connect the cDAQ-9178 chassis to your host computer.

6.

Connect the power source to the cDAQ-9178 chassis.

7.

Launch Measurement & Automation Explorer (MAX).

8.

Right-click the device name and select 

Self-Test

 to ensure that the module is working 

properly.

Содержание TB-9212

Страница 1: ...sively referred to as TB 9212 For more information about calibration solutions visit ni com calibration Note Refer to the NI 9212 Calibration Procedure on ni com manuals for instructions on calibratin...

Страница 2: ...com and help files install with the software NI cDAQ 9174 9178 Quick Start NI DAQmx installation and hardware setup NI 9212 Getting Started Guide NI 9212 with TB 9212 specific information NI 9212 Dat...

Страница 3: ...212 An NI 9212 module that is verified to be within product specifications Chassis cDAQ 9178 Barrier Strip Assembly Thermocouple Wire TT E 24 SLE 100 E type SLE 24 AWG 20 m to 30 m Mini Thermocouple C...

Страница 4: ...re assembly Crimp the chromel spade lugs to the chromel wire to create 8 positive spade lugs Crimp the constantan spade lugs to the constantan wire to create 8 negative spade lugs 2 Use the thermocoup...

Страница 5: ...how to install the software and hardware and how to configure the device in Measurement Automation Explorer MAX Complete the following steps to set up the TB 9212 1 Install NI DAQmx Note Ensure that...

Страница 6: ...ou are using a TB 9212 with mini TC you do not need to remove the top cover 2 Connect the thermocouple calibrator to the TB 9212 using the barrier strip assembly created in the Building the Barrier St...

Страница 7: ...eliminates gain errors in the NI 9212 9 Acquire and average samples a Create and configure an AI temperature thermocouple channel for the NI 9212 according to Table 4 b Configure the AI temperature th...

Страница 8: ...Repeat step 9 12 Perform the following calculation using the recorded Tempaix and CJCtemp x values TempErrorCh0 3 Tempai0 3 CJCtemp0 TempErrorCh4 7 Tempai4 7 CJCtemp1 13 Compare the calculation resul...

Страница 9: ...complete resource for technical support At ni com support you have access to everything from troubleshooting and application development self help resources to email and phone assistance from NI Appli...

Отзывы: