National Instruments Corporation
9
SCXI-1520 Calibration Procedure
Adjusting for Gain, Offset, and Excitation Errors
The
and
Adjusting Calibration Constants
sections explain how to determine the gain, offset, and excitation errors of
the SCXI-1520 and how to adjust the calibration constants to correct for
these errors.
Determining Gain Errors
Complete the following steps to determine the gain errors of the
SCXI-1520.
1.
Read the
2.
Refer to the specifications in Table 2.
3.
Call
Calibrate_E_Series
to minimize the uncertainty associated
with the DAQ device measurements. Set the following parameters:
•
deviceNumber
—The value assigned by MAX
•
calOP
—
ND_SELF_CALIBRATE
•
setOfCalConst
—
ND_USER_EEPROM_AREA
•
calRefVolts
—0.0
4.
Call
AI_Configure
to put the DAQ device in bipolar mode. Set the
following parameters:
•
deviceNumber
—The value assigned by MAX
•
chan
—0
•
inputMode
—0
•
inputRange
—0
•
polarity
—0
•
driveAIS
—0
5.
Call
SCXI_Track_Hold_Setup
to configure the track-and-hold
circuitry of the SCXI-1520. Set the following parameters:
•
SCXIchassisID
—The value assigned by MAX
•
moduleSlot
—1
•
inputMode
—2
•
source
—0
•
send
—0
•
holdCount
—1
•
DAQdeviceNumber
—The value assigned by MAX