SCXI-1520 Calibration Procedure
2
ni.com
•
NI-DAQ User Manual for PC Compatibles
•
SCXI-1314 Installation Guide
•
SCXI-1349 Installation Guide
Software
The SCXI-1520 calibration procedure requires the latest version of the
NI-DAQ driver on the calibration system. The driver supports a number
of programming languages, including LabVIEW, Measurement Studio,
Microsoft Visual C++, Microsoft Visual Basic, and Borland C++. When
you install the driver, you need to install support only for the programming
language you are using.
Documentation
The
NI-DAQ Function Reference Help
and the
NI-DAQ User Manual for
PC Compatibles
contain detailed information on using the NI-DAQ driver.
The online help includes detailed information on the driver functions.
You can access the help by selecting
Start»Programs»National
Instruments DAQ»NI-DAQ Help
. The user manual provides instructions
on installing and configuring DAQ devices. The user manual also includes
detailed information on creating applications that use the NI-DAQ driver.
These are your primary references for writing your calibration utility. For
more information on the SCXI-1520, refer to the
SCXI-1520 User Manual
.
Test Conditions
Follow these guidelines to optimize the connections and the environment
during calibration.
•
Keep connections to the SCXI module short. Long cables and wires act
as antennae, picking up extra noise and thermal offsets that can affect
measurements.
•
Use shielded copper wire for all cable connections to the device.
Use twisted-pair wire to reduce noise.
•
Maintain the temperature between 18 and 28 °C.
•
Keep relative humidity below 80%.
•
Allow a warm-up time of at least 15 minutes for the SCXI module
and 30 minutes for the DAQ device to ensure that the measurement
circuitry is at a stable operating temperature.