background image

©

 National Instruments Corporation

7

SCXI-1141/1142/1143 Calibration Procedure

moduleSlot

—1

channel

—0

filterMode

—0

freq

—0 kHz

cutoffDivDown

—0

outClkDivDown

—2

9.

Input the test voltage listed in Table 9 to channel 0 of the SCXI 
module.

10. Call 

DAQ_Op

. Set the following parameters:

DAQdeviceNumber

—The device number for the DAQ device

channel

—0

gain

—1 for a 16-bit E Series DAQ device

count

—100

sampleRate

—100

11. Call 

SCXI_Scale

 to convert the readings from binary to voltage. 

Set the following parameters:

SCXIChassisID

—The device number assigned by MAX

moduleSlot

—1

channel

—0

SCXIgain

—The gain you are testing

TBgain

—1

DAQboard

—The device number for the DAQ device

DAQChannel

—0

DAQgain

—1

numPoints

—100

binArray

—The array returned from 

DAQ_Op

The result is a set of scaled voltages read from the SCXI module.

12. Average the results returned by 

SCXI_Scale

. Compare the averaged 

result to the upper and lower limits listed in Table 9.

13. Repeat steps 6 through 12 for each remaining test point.

14. Repeat steps 6 through 13 for each remaining channel, changing the 

moduleChan 

variable to the channel number you are testing.

You have finished verifying the operation of the SCXI-1141/1142/1143.

Содержание SCXI-1141

Страница 1: ...e sequence File Page Setup Options directs you to pull down the File menu select the Page Setup item and select Options from the last dialog box This icon denotes a tip which alerts you to advisory in...

Страница 2: ...acy which includes high level function calls to simplify the task of writing software to calibrate devices Traditional NI DAQ Legacy supports many programming languages including LabVIEW LabWindows CV...

Страница 3: ...install Traditional NI DAQ Legacy After adding these help files you can access them by selecting Start Programs National Instruments NI DAQ Online DAQ Support The documents provide instructions about...

Страница 4: ...calling the appropriate calibration functions The steps used in the calibration procedure are as follows 1 Setting up the SCXI 1141 1142 1143 for testing 2 Verifying the existing operation of the SCXI...

Страница 5: ...w well the SCXI 1141 1142 1143 is meeting its specifications You can use this information to select the appropriate calibration interval for your application 1 SCXI 1304 Terminal Block 2 SCXI 1141 114...

Страница 6: ...d by MAX calOp ND_SELF_CALIBRATE setOfCalConst ND_USER_EEPROM_AREA calRefVolts 0 0 5 Call MIO_Config to enable dithering on the E Series DAQ device measurements Set the following parameters DAQdeviceN...

Страница 7: ...arameters SCXIChassisID The device number assigned by MAX moduleSlot 1 channel 0 SCXIgain The gain you are testing TBgain 1 DAQboard The device number for the DAQ device DAQChannel 0 DAQgain 1 numPoin...

Страница 8: ...g input channel 0 Refer to Table 7 to determine which pins on the 96 pin front connector correspond to the positive and negative inputs of the specified channel If you are using an SCXI 1304 connected...

Страница 9: ...et the calibrator output to a 1 Vrms sine wave at the frequency specified in Tables 4 5 or 6 2 Verify that the DMM output is within the bounds specified by Tables 4 5 or 6 3 Repeat steps 1 and 2 until...

Страница 10: ...he SCXI 1141 1142 1143 4 Call SCXI_Single_Chan_Setup to configure the module for single channel measurements Set the following parameters SCXIchassisID The value obtained from MAX moduleSlot 1 moduleC...

Страница 11: ...for later use 10 Set the calibrator to the negative test point value for the same gain Skip any input limits that are specified as 0 0 V You need only upper and lower limits for adjustment 11 Read th...

Страница 12: ...slot channel The channel you want to adjust opCode 2 calibrationArea 0 rangeCode 0 not used for the SCXI 1141 1142 1143 SCXIgain The gain setting you want to adjust DAQboard The device number assigned...

Страница 13: ...the sine wave must be lower than 1 50th of the cutoff frequency For this procedure in sections SCXI 1141 1142 1143 Before Revision F or SCXI 1141 1142 1143 Revision F or Later you will set the filters...

Страница 14: ...E Series DAQ device is connected to the SCXI 1141 1142 1143 7 Call SCXI_Set_Gain to configure the module to a gain value of 1 Set the following parameters SCXIchassisID The device number assigned by M...

Страница 15: ...channel 0 is pin 4 which is AI 0 11 Set the calibrator voltage to 3 4 Vrms 400 Hz 12 Measure and record the amplitude of the sine wave with the DMM 13 Call SCXI_Configure_Filter to disable the filter...

Страница 16: ...nputs of the specified channel If you are using an SCXI 1304 that is connected to the SCXI 1141 1142 1143 connect the calibrator to the AI 0 and AI 0 inputs 6 Connect the DMM to the output of channel...

Страница 17: ...MAX moduleSlot 4 channel 0 value The new value decided upon in step 12 14 Repeat steps 11 through 13 until the measured amplitude is as close as possible to the level it was with the filter in bypass...

Страница 18: ...3 NC AI 3 23 NC NC NC 22 A GND NC A GND 21 NC NC NC 20 AI 4 NC AI 4 19 NC NC NC 18 AI 5 NC AI 5 17 NC NC NC 16 A GND NC A GND 15 NC NC NC 14 AI 6 NC AI 6 13 NC NC NC 12 AI 7 NC AI 7 11 NC NC NC 10 NC...

Страница 19: ...GND AI 3 9 10 A GND AI 4 11 12 A GND AI 5 13 14 A GND AI 6 15 16 A GND AI 7 17 18 A GND NC 19 20 NC NC 21 22 NC NC 23 24 DIG GND SER DAT IN 25 26 SER DAT OUT DAQ D A 27 28 NC SLOT 0 SEL 29 30 NC NC 31...

Страница 20: ...41 1142 1143 If the module has been calibrated within the last year the Test Point V value should fall between the Upper Limit V and Lower Limit V values Table 9 SCXI 1141 1142 1143 Specifications Tes...

Отзывы: