background image

©

 National Instruments Corporation

11

SCXI-1141/1142/1143 Calibration Procedure

6.

Connect the calibrator to the appropriate analog input channel, starting 
with channel 0. Refer to Table 7 to determine the pins on the 96-pin 
front connector that correspond to the positive and negative inputs of 
the specified channel. If you are using an SCXI-1304 connected to the 
SCXI-1141/1142/1143, connect the calibrator to the AI 0 + and AI 0 – 
inputs and verify that the SCXI-1304 is set to DC coupling.

7.

Connect the DMM to the positive output on pin 3 (AI 0 +) and the 
negative output on pin 4 (AI 0 –) on the rear panel connector, 
illustrated in Table 8.

Tip

For easy access to individual pins, use a TBX 50-pin connector block connected as 

shown in Figure 1.

8.

Set the calibrator voltage to the positive test-point value in Table 9.

9.

Read the voltage from the DMM. Record the DMM reading as 

output1

, and the calibrator output voltage as volt1, for later use.

10. Set the calibrator to the negative test-point value for the same gain. 

Skip any input limits that are specified as 0.0 V. You need only upper 
and lower limits for adjustment.

11. Read the voltage from the DMM. Record the DMM reading as 

output2

, and the calibrator output voltage as 

volt2

. You now have two 

pairs of data points, (

volt1

output1

) and (

volt2

output2

), where 

volt1

 

and 

volt2

 are calibrator readings, and 

output1

 and 

output2

 are DMM 

readings.

12. Convert the DMM voltage readings (

output1

 and 

output2

) to binary 

readings (

binary1

 and 

binary2

) using the following equation:

Note

In both of the above equations, 

output

 is either 

output1

 or 

output2

.

For example, using a 16-bit product such as a NI 6030E and obtaining 
a DMM reading of –9.90000 V, you get the following results:

13. Record 

binary1

 with 

volt1

, and 

binary2

 with 

volt2

, for later use.

14. Repeat steps 5 through 13 for the remaining gain values you want to 

measure on this channel.

15. Repeat steps 4 through 14 for the remaining channels you want to 

measure.

binary reading for 16-bit E Series products (binary) 

output

20

-----------------

2

16

×

=

binary

9.90000

20

----------------------

2

16

×

32440.32

=

=

Содержание SCXI-1141

Страница 1: ...e sequence File Page Setup Options directs you to pull down the File menu select the Page Setup item and select Options from the last dialog box This icon denotes a tip which alerts you to advisory in...

Страница 2: ...acy which includes high level function calls to simplify the task of writing software to calibrate devices Traditional NI DAQ Legacy supports many programming languages including LabVIEW LabWindows CV...

Страница 3: ...install Traditional NI DAQ Legacy After adding these help files you can access them by selecting Start Programs National Instruments NI DAQ Online DAQ Support The documents provide instructions about...

Страница 4: ...calling the appropriate calibration functions The steps used in the calibration procedure are as follows 1 Setting up the SCXI 1141 1142 1143 for testing 2 Verifying the existing operation of the SCXI...

Страница 5: ...w well the SCXI 1141 1142 1143 is meeting its specifications You can use this information to select the appropriate calibration interval for your application 1 SCXI 1304 Terminal Block 2 SCXI 1141 114...

Страница 6: ...d by MAX calOp ND_SELF_CALIBRATE setOfCalConst ND_USER_EEPROM_AREA calRefVolts 0 0 5 Call MIO_Config to enable dithering on the E Series DAQ device measurements Set the following parameters DAQdeviceN...

Страница 7: ...arameters SCXIChassisID The device number assigned by MAX moduleSlot 1 channel 0 SCXIgain The gain you are testing TBgain 1 DAQboard The device number for the DAQ device DAQChannel 0 DAQgain 1 numPoin...

Страница 8: ...g input channel 0 Refer to Table 7 to determine which pins on the 96 pin front connector correspond to the positive and negative inputs of the specified channel If you are using an SCXI 1304 connected...

Страница 9: ...et the calibrator output to a 1 Vrms sine wave at the frequency specified in Tables 4 5 or 6 2 Verify that the DMM output is within the bounds specified by Tables 4 5 or 6 3 Repeat steps 1 and 2 until...

Страница 10: ...he SCXI 1141 1142 1143 4 Call SCXI_Single_Chan_Setup to configure the module for single channel measurements Set the following parameters SCXIchassisID The value obtained from MAX moduleSlot 1 moduleC...

Страница 11: ...for later use 10 Set the calibrator to the negative test point value for the same gain Skip any input limits that are specified as 0 0 V You need only upper and lower limits for adjustment 11 Read th...

Страница 12: ...slot channel The channel you want to adjust opCode 2 calibrationArea 0 rangeCode 0 not used for the SCXI 1141 1142 1143 SCXIgain The gain setting you want to adjust DAQboard The device number assigned...

Страница 13: ...the sine wave must be lower than 1 50th of the cutoff frequency For this procedure in sections SCXI 1141 1142 1143 Before Revision F or SCXI 1141 1142 1143 Revision F or Later you will set the filters...

Страница 14: ...E Series DAQ device is connected to the SCXI 1141 1142 1143 7 Call SCXI_Set_Gain to configure the module to a gain value of 1 Set the following parameters SCXIchassisID The device number assigned by M...

Страница 15: ...channel 0 is pin 4 which is AI 0 11 Set the calibrator voltage to 3 4 Vrms 400 Hz 12 Measure and record the amplitude of the sine wave with the DMM 13 Call SCXI_Configure_Filter to disable the filter...

Страница 16: ...nputs of the specified channel If you are using an SCXI 1304 that is connected to the SCXI 1141 1142 1143 connect the calibrator to the AI 0 and AI 0 inputs 6 Connect the DMM to the output of channel...

Страница 17: ...MAX moduleSlot 4 channel 0 value The new value decided upon in step 12 14 Repeat steps 11 through 13 until the measured amplitude is as close as possible to the level it was with the filter in bypass...

Страница 18: ...3 NC AI 3 23 NC NC NC 22 A GND NC A GND 21 NC NC NC 20 AI 4 NC AI 4 19 NC NC NC 18 AI 5 NC AI 5 17 NC NC NC 16 A GND NC A GND 15 NC NC NC 14 AI 6 NC AI 6 13 NC NC NC 12 AI 7 NC AI 7 11 NC NC NC 10 NC...

Страница 19: ...GND AI 3 9 10 A GND AI 4 11 12 A GND AI 5 13 14 A GND AI 6 15 16 A GND AI 7 17 18 A GND NC 19 20 NC NC 21 22 NC NC 23 24 DIG GND SER DAT IN 25 26 SER DAT OUT DAQ D A 27 28 NC SLOT 0 SEL 29 30 NC NC 31...

Страница 20: ...41 1142 1143 If the module has been calibrated within the last year the Test Point V value should fall between the Upper Limit V and Lower Limit V values Table 9 SCXI 1141 1142 1143 Specifications Tes...