Chapter 3
Configuring Your SCXI Hardware and Software
©
National Instruments Corporation
3-11
2.
Configure the filtering.
♦
SCXI-1120 Filter Jumpers
Two-stage filtering is also available on your SCXI-1120 module. The
first stage is located in the isolated section of the input channel,
whereas the second stage is located in the nonisolated section of your
input channel. Two-stage filtering eliminates the noise generated by
the isolation amplifier, producing a higher signal-to-noise ratio.
Furthermore, two filter bandwidths are available, 10 kHz and 4 Hz.
Second-stage
1
A
2
B
5
C
10
D (factory-default
setting for the
SCXI-1120)
20
E (factory-default
setting for the
SCXI-1120D)
Table 3-6.
SCXI-1120 Filter Jumper Allocation
Input
Channel
Number
First
Filter Jumper
Second
Filter Jumper
4 Hz
(Factory
Default)
10 kHz
4 Hz
(Factory
Default)
10 kHz
0
W17-A
W17-B
W25
W26
1
W18-A
W18-B
W27
W28
2
W19-A
W19-B
W29
W30
3
W20-A
W20-B
W31
W32
Table 3-5.
Gain Jumper Positions (Continued)
Gain
Setting
Jumper Position