Using the SCC-LP
XX
when Scaling Voltage Measurements
If you configured the SCC-LP
XX
using Measurement & Automation
Explorer (MAX) and you are using NI-DAQ, the voltage reading you get
from the E Series DAQ device accounts for the voltage scaling effect of the
SCC-LP
XX
. Otherwise, you must scale the readings as follows:
1.
Read the SCC-LP
XX
channel on the E Series DAQ device
V
ESERIES
(CH
X
).
2.
Calculate the SCC-LP
XX
voltage using this formula:
V
LP
= 2
V
ESERIES
where
V
LP
is the SCC-LP
XX
input voltage.
V
ESERIES
is the E Series DAQ device input voltage.
Specifications
These ratings are typical at 25 °C unless otherwise stated.
Amplifier Characteristics
Number of input channels.......................2 DIFF
Input signal range
...................................±
10 V
Output signal range
.................................±
5 V
Gain ........................................................0.5
Input impedance .....................................10 G
Ω
in parallel with 10 pF
(powered on)
10 k
Ω
(powered off or overloaded)
Gain error................................................Adjustable to 0% of reading
Offset-voltage error ................................350
µ
V typ
(referred to input, RTI)
1
1.5 mV max
2
1
This specification is calculated relative to the input range of the module.
2